skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Absolute calibration in grazing incidence interferometry via rotational averaging

Journal Article · · Applied Optics
DOI:https://doi.org/10.1364/AO.45.0· OSTI ID:20779345

Interferometry in grazing incidence can be used to test cylindrical mantle surfaces. The absolute accuracy of the resulting surface profiles is limited by systematic wavefront aberrations caused in the interferometer, in particular due to an inversion of the test wavefront in an interferometer using diffractive beam splitters. For cylindrical specimens, a calibration method using four positions has therefore been investigated. This test is combined with another method of optical metrology: the rotational averaging procedure. The implementation for grazing incidence is described and measurement results for hollow cylinders are presented. The gain in accuracy is demonstrated.

OSTI ID:
20779345
Journal Information:
Applied Optics, Vol. 45, Issue 16; Other Information: DOI: 10.1364/AO.45.003740; (c) 2006 Optical Society of America; Country of input: International Atomic Energy Agency (IAEA); ISSN 0003-6935
Country of Publication:
United States
Language:
English