skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Design and operational characteristics of a cast steel mass spectrometer

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1784564· OSTI ID:20643963
; ;  [1]
  1. National Institute of Physics, College of Science, University of the Philippines, Diliman, Quezon City 1101 (Philippines)

A cast steel magnetic sector mass analyzer is developed for studies of hydrogen and helium ion beams generated by a gas discharge compact ion source. The optimum induced magnetic flux density of 3500 G made it possible to scan the whole spectrum of hydrogen and helium ion species. Analysis of beam characteristics shows that the mass spectrometer sensitivity, and resolving power are approximately inversely proportional. The resolution is enhanced at higher pressures and lower current discharges. In contrast, the instrument sensitivity increased at higher current discharges and decreased at higher pressures. Calculations of the ultimate resolving power with reference to analyzer dimensions yield a numerical value of 30. System anomaly in the form of spherical aberrations was also analyzed using the paraxial beam envelope equation. Beam divergence is most significant at high discharge conditions where angular spread reaches an upper limit of 8.6 deg.

OSTI ID:
20643963
Journal Information:
Review of Scientific Instruments, Vol. 75, Issue 9; Other Information: DOI: 10.1063/1.1784564; (c) 2004 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English