A simple method of independent treatment time verification in gamma knife radiosurgery using integral dose
- Department of Radiation Oncology, Siteman Cancer Center, Washington University Medical Center, St. Louis, Missouri 63110 (United States)
The purpose of this study is to develop a simple independent dose calculation method to verify treatment plans for Leksell Gamma Knife radiosurgery. Our approach uses the total integral dose within the skull as an end point for comparison. The total integral dose is computed using a spreadsheet and is compared to that obtained from Leksell GammaPlan registered . It is calculated as the sum of the integral doses of 201 beams, each passing through a cylindrical volume. The average length of the cylinders is estimated from the Skull-Scaler measurement data taken before treatment. Correction factors are applied to the length of the cylinder depending on the location of a shot in the skull. The radius of the cylinder corresponds to the collimator aperture of the helmet, with a correction factor for the beam penumbra and scattering. We have tested our simple spreadsheet program using treatment plans of 40 patients treated with Gamma Knife registered in our center. These patients differ in geometry, size, lesion locations, collimator helmet, and treatment complexities. Results show that differences between our calculations and treatment planning results are typically within {+-}3%, with a maximum difference of {+-}3.8%. We demonstrate that our spreadsheet program is a convenient and effective independent method to verify treatment planning irradiation times prior to implementation of Gamma Knife radiosurgery.
- OSTI ID:
- 20634526
- Journal Information:
- Medical Physics, Vol. 31, Issue 12; Other Information: DOI: 10.1118/1.1824192; (c) 2004 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
WE-G-BRA-08: Failure Modes and Effects Analysis (FMEA) for Gamma Knife Radiosurgery
Air kerma based dosimetry calibration for the Leksell Gamma Knife