Experimental results and experience with the KfK-tritium-calorimeter
- Forschungszentrum Karlsruhe (Germany)
For the tritium assay at the Tritium Laboratory Karlsruhe (TLK) which started operation in Nov.94 an isothermal shield calorimeter was built. The calibration of the calorimeter was performed with an electrically heated simulator covering the range between 1 mW and about 2 W. After the calibration tritium activities in four Amersham uranium getter containers were determined. The excellent reproducibility and high precision of the measurements with the KfK-calorimeter was shown. The tritium activity inventories could be determined with relative standard deviations from the average value of the order of 0.01% to 0.1% depending on the activity level. These results demonstrate the suitability of the KfK-calorimeter as a nondestructive method for a careful assay of tritium. 3 refs., 11 figs., 2 tabs.
- OSTI ID:
- 196743
- Report Number(s):
- CONF-950506-; ISSN 0748-1896; TRN: 96:001398-083
- Journal Information:
- Fusion Technology, Vol. 28, Issue 3 pt 2; Conference: 5. topical meeting on tritium technology in fission, fusion and isotopic applications, Ispra (Italy), 28 May - 3 Jun 1995; Other Information: PBD: Oct 1995
- Country of Publication:
- United States
- Language:
- English
Similar Records
Tritium and uranium inventory measurements with the JET AGHS precision calorimeter
Current status of tritium calorimetry at TLK
Related Subjects
05 NUCLEAR FUELS
99 MATHEMATICS
COMPUTERS
INFORMATION SCIENCE
MANAGEMENT
LAW
MISCELLANEOUS
TRITIUM
INVENTORIES
ACCOUNTING
CALORIMETERS
SIMULATION
STATISTICAL MECHANICS
NONDESTRUCTIVE TESTING
CALIBRATION
TEMPERATURE DISTRIBUTION
ALGORITHMS
THERMONUCLEAR FUELS
NUMERICAL DATA