skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Experimental results and experience with the KfK-tritium-calorimeter

Journal Article · · Fusion Technology
OSTI ID:196743
;  [1]
  1. Forschungszentrum Karlsruhe (Germany)

For the tritium assay at the Tritium Laboratory Karlsruhe (TLK) which started operation in Nov.94 an isothermal shield calorimeter was built. The calibration of the calorimeter was performed with an electrically heated simulator covering the range between 1 mW and about 2 W. After the calibration tritium activities in four Amersham uranium getter containers were determined. The excellent reproducibility and high precision of the measurements with the KfK-calorimeter was shown. The tritium activity inventories could be determined with relative standard deviations from the average value of the order of 0.01% to 0.1% depending on the activity level. These results demonstrate the suitability of the KfK-calorimeter as a nondestructive method for a careful assay of tritium. 3 refs., 11 figs., 2 tabs.

OSTI ID:
196743
Report Number(s):
CONF-950506-; ISSN 0748-1896; TRN: 96:001398-083
Journal Information:
Fusion Technology, Vol. 28, Issue 3 pt 2; Conference: 5. topical meeting on tritium technology in fission, fusion and isotopic applications, Ispra (Italy), 28 May - 3 Jun 1995; Other Information: PBD: Oct 1995
Country of Publication:
United States
Language:
English