Radioxenon detector calibration spike production and delivery systems
Abstract Beta-Gamma coincidence radioxenon detectors must be calibrated for each of the four-radioxenon isotopes (135Xe, 133Xe, 133mXe, and 131mXe). Without a proper calibration, there is potential for the misidentification of the amount of each isotope detected. It is important to accurately determine the amount of each radioxenon isotope, as the ratios can be used to distinguish between an anthropogenic source and a nuclear explosion. We have developed a xenon calibration system (XeCalS) that produces calibration spikes of known activity and pressure for field calibration of detectors. The activity concentrations of these calibration spikes are measured using a beta-gamma coincidence detector and a high purity germanium (HPGe) detector. We will present the results from the development and commissioning of XeCalS, along with the future plans for a portable spike implementation system.