Title: Improved precision and accuracy in quantifying plutonium isotope ratios by RIMS

Journal Article · · Journal of Radioanalytical and Nuclear Chemistry
 [1];  [2];  [1];  [1];  [1];  [1]
  1. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
  2. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Argonne National Lab. (ANL), Argonne, IL (United States)

Resonance ionization mass spectrometry (RIMS) holds the promise of rapid, isobar-free quantification of actinide isotope ratios in as-received materials (i.e. not chemically purified). Recent progress in achieving this potential using two Pu test materials is presented. RIMS measurements were conducted multiple times over a period of two months on two different Pu solutions deposited on metal surfaces. Measurements were bracketed with a Pu isotopic standard, and yielded absolute accuracies of the measured 240Pu/239Pu ratios of 0.7% and 0.58%, with precisions (95% confidence intervals) of 1.49% and 0.91%. In conclusion, the minor isotope 238Pu was also quantified despite the presence of a significant quantity of 238U in the samples.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Basic Energy Sciences (BES)
Grant/Contract Number:
AC52-07NA27344; AC02-06CH11357
OSTI ID:
1241996
Report Number(s):
LLNL-JRNL-674877; TRN: US1600407
Journal Information:
Journal of Radioanalytical and Nuclear Chemistry, Vol. 307, Issue 3; ISSN 0236-5731
Publisher:
SpringerCopyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 8 works
Citation information provided by
Web of Science

References (17)

Isotopic analysis of uranium and plutonium mixtures by resonance ionization mass spectrometry journal March 1984
Measurement of plutonium isotope ratios in nuclear fuel samples by HPLC-MC-ICP-MS journal April 2005
Detection of plutonium isotopes at lowest quantities using in-source resonance ionization mass spectrometry journal August 2012
Improvements in RIMS Isotopic Precision: Application to in situ atom-limited analyses conference January 2009
Resonance ionization mass spectrometry for precise measurements of isotope ratios journal November 2009
Improving Precision in Resonance Ionization Mass Spectrometry: Influence of Laser Bandwidth in Uranium Isotope Ratio Measurements journal April 2011
Determination of fission products in nuclear samples by capillary electrophoresis-inductively coupled plasma mass spectrometry (CE-ICP-MS) journal March 2008
Production and Characterization of Monodisperse Plutonium, Uranium, and Mixed Uranium−Plutonium Particles for Nuclear Safeguard Applications journal May 2010
Critical comparison of radiometric and mass spectrometric methods for the determination of radionuclides in environmental, biological and nuclear waste samples journal February 2008
Analyzing individual presolar grains with CHARISMA journal September 2003
Laser-induced isotopic selectivity in the resonance ionization of Os journal June 1992
Correction of dead time effects in laser-induced desorption time-of-flight mass spectrometry: Applications in atom probe tomography journal March 2015
Improved isotopic SIMS measurements of uranium particles for nuclear safeguard purposes journal January 2009
Resonance ionization mass spectrometry for ultratrace analysis of plutonium with a new solid state laser system journal July 2004
Correction of dead time effects in time‐of‐flight mass spectrometry
  • Stephan, T.; Zehnpfenning, J.; Benninghoven, A.
  • Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films, Vol. 12, Issue 2 https://doi.org/10.1116/1.579255
journal March 1994
Laser-induced isotopic effects in titanium resonance ionization journal May 1993
Ultratrace analysis and isotope ratio measurements of long-lived radioisotopes by resonance ionization mass spectrometry (RIMS) journal January 2004

Cited By (5)

Analysis of sub-picogram quantities of 238Pu by thermal ionization mass spectrometry journal July 2019
Atomic spectrometry update: review of advances in the analysis of metals, chemicals and materials journal January 2017
Atomic spectrometry update: review of advances in the analysis of metals, chemicals and materials journal January 2018
Atomic spectrometry update: review of advances in the analysis of metals, chemicals and materials journal January 2020
Atomic spectrometry update. Review of advances in the analysis of metals, chemicals and materials journal January 2013