skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: ToF-SIMS Depth Profiling Of Insulating Samples, Interlaced Mode Or Non-interlaced Mode?

Journal Article · · Surface and Interface Analysis, 46(S1):257-260
DOI:https://doi.org/10.1002/sia.5419· OSTI ID:1176850

Dual beam depth profiling strategy has been widely adopted in ToF-SIMS depth profiling, in which two basic operation modes, interlaced mode and non-interlaced mode, are commonly used. Generally, interlaced mode is recommended for conductive or semi-conductive samples, whereas non-interlaced mode is recommended for insulating samples, where charge compensation can be an issue. Recent publications, however, show that the interlaced mode can be used effectively for glass depth profiling, despite the fact that glass is an insulator. In this study, we provide a simple guide for choosing between interlaced mode and non-interlaced mode for insulator depth profiling. Two representative cases are presented: (1) depth profiling of a leached glass sample, and (2) depth profiling of a single crystal MgO sample. In brief, the interlaced mode should be attempted first, because (1) it may provide reasonable-quality data, and (2) it is time-saving for most cases, and (3) it introduces low H/C/O background. If data quality is the top priority and measurement time is flexible, non-interlaced mode is recommended because interlaced mode may suffer from low signal intensity and poor mass resolution. A big challenge is tracking trace H/C/O in a highly insulating sample (e.g., MgO), because non-interlaced mode may introduce strong H/C/O background but interlaced mode may suffer from low signal intensity. Meanwhile, a C or Au coating is found to be very effective to improve the signal intensity. Surprisingly, the best analyzing location is not on the C or Au coating, but at the edge (outside) of the coating.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Environmental Molecular Sciences Lab. (EMSL)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
1176850
Report Number(s):
PNNL-SA-99490; 47459; 47580; KP1704020
Journal Information:
Surface and Interface Analysis, 46(S1):257-260, Journal Name: Surface and Interface Analysis, 46(S1):257-260
Country of Publication:
United States
Language:
English

Similar Records

An Investigation of Hydrogen Depth Profiling Using ToF-SIMS
Journal Article · Wed Feb 01 00:00:00 EST 2012 · Surface and Interface Analysis · OSTI ID:1176850

Nanoscale imaging of hydrogen and sodium in alteration layers of corroded glass using ToF-SIMS: Is an auxiliary sputtering ion beam necessary?
Journal Article · Thu Oct 11 00:00:00 EDT 2018 · Surface and Interface Analysis · OSTI ID:1176850

Ultra-shallow junction depth profile analysis using TOF-SIMS and TXRF
Journal Article · Tue Nov 24 00:00:00 EST 1998 · AIP Conference Proceedings · OSTI ID:1176850