skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Studies of single-photoelectron response and of performance in magnetic field of a H8500C-03 photomultiplier tube

Journal Article · · Journal of Instrumentation

We studied the single-photoelectron detection capabilities of a multianode photomultiplier tube H8500C-03 and its performance in high magnetic field. Our results show that the device can readily resolve signals at the single photoelectron level making it suitable for photon detection in both threshold and ring imaging Cherenkov detectors. We also found that a large longitudinal magnetic field, up to 300 Gauss, induces a change in the relative output of at most 55% for an edge pixel, and of at most 15% for a central pixel. The H8500C-03 signal loss in transverse magnetic fields it is significantly more pronounced than for the longitudinal case. Our studies of single photoelectron reduction in magnetic fields point to the field induced misfocusing of the photoelectron extracted from the photocathode as primary cause of signal loss. With appropriate shielding this PMT could function in high magnetic field environments.

Research Organization:
Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
AC05-06OR23177; FG02-03ER41231
OSTI ID:
1108714
Report Number(s):
JLAB-PHY-13-1819; DOE/OR/23177-2933
Journal Information:
Journal of Instrumentation, Vol. 8
Country of Publication:
United States
Language:
English

Similar Records

Characterization of Multianode Photomultiplier Tubes for use in the CLAS12 RICH detector
Journal Article · Mon Sep 19 00:00:00 EDT 2022 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1108714

Replacing photomultiplier tubes with silicon photomultipliers for nuclear safeguards applications
Conference · Tue Dec 01 00:00:00 EST 2020 · Proceedings of the INMM 61st Annual Meeting · OSTI ID:1108714

Replacing photomultiplier tubes with silicon photomultipliers for nuclear safeguards applications
Technical Report · Mon Oct 12 00:00:00 EDT 2020 · OSTI ID:1108714