Title: Alignment of a vector magnetometer to an optical prism

Abstract

A method for alignment of a vector magnetometer to a rigidly attached prism is presented. This enables optical comparison of the magnetometer axes to physical surfaces in the apparatus, and thus an absolute determination of the magnetic field direction in space. This is in contrast with more common techniques, which focus on precise determination of the relative angles between magnetometer axes, and so are more suited to measuring differences in the direction of magnetic fields. Here we demonstrate precision better than 500 μrad on a fluxgate magnetometer, which also gives the coil orthogonality errors to a similar precision. As a result, the relative sensitivity of the three axes is also determined, with a precision of about 5 ×10–4.

Authors:
 [1];  [1];  [1]
  1. Argonne National Lab. (ANL), Lemont, IL (United States)
Publication Date:
Research Org.:
Argonne National Lab. (ANL), Argonne, IL (United States)
Sponsoring Org.:
USDOE Office of Science (SC), Nuclear Physics (NP)
OSTI Identifier:
1365396
Alternate Identifier(s):
OSTI ID: 1361882
Grant/Contract Number:  
AC02-06CH11357
Resource Type:
Accepted Manuscript
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 88; Journal Issue: 5; Journal ID: ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY

Citation Formats

Dietrich, M. R., Bailey, K. G., and O’Connor, T. P. Alignment of a vector magnetometer to an optical prism. United States: N. p., 2017. Web. doi:10.1063/1.4983146.
Dietrich, M. R., Bailey, K. G., & O’Connor, T. P. Alignment of a vector magnetometer to an optical prism. United States. https://doi.org/10.1063/1.4983146
Dietrich, M. R., Bailey, K. G., and O’Connor, T. P. Fri . "Alignment of a vector magnetometer to an optical prism". United States. https://doi.org/10.1063/1.4983146. https://www.osti.gov/servlets/purl/1365396.
@article{osti_1365396,
title = {Alignment of a vector magnetometer to an optical prism},
author = {Dietrich, M. R. and Bailey, K. G. and O’Connor, T. P.},
abstractNote = {A method for alignment of a vector magnetometer to a rigidly attached prism is presented. This enables optical comparison of the magnetometer axes to physical surfaces in the apparatus, and thus an absolute determination of the magnetic field direction in space. This is in contrast with more common techniques, which focus on precise determination of the relative angles between magnetometer axes, and so are more suited to measuring differences in the direction of magnetic fields. Here we demonstrate precision better than 500 μrad on a fluxgate magnetometer, which also gives the coil orthogonality errors to a similar precision. As a result, the relative sensitivity of the three axes is also determined, with a precision of about 5 ×10–4.},
doi = {10.1063/1.4983146},
journal = {Review of Scientific Instruments},
number = 5,
volume = 88,
place = {United States},
year = {Fri May 12 00:00:00 EDT 2017},
month = {Fri May 12 00:00:00 EDT 2017}
}

Journal Article:
Free Publicly Available Full Text
Publisher's Version of Record

Citation Metrics:
Cited by: 3 works
Citation information provided by
Web of Science

Save / Share:

Works referenced in this record:

A Procedure for Accurate Calibration of the Orientation of the Three Sensors in a Vector Magnetometer
journal, April 1978

  • McPherron, Robert; Snare, Robert
  • IEEE Transactions on Geoscience Electronics, Vol. 16, Issue 2
  • DOI: 10.1109/tge.1978.294576

Magnetic Field Uniformity around Near‐Helmholtz Coil Configurations
journal, November 1969

  • Cacak, R. K.; Craig, J. R.
  • Review of Scientific Instruments, Vol. 40, Issue 11
  • DOI: 10.1063/1.1683829

Calibration of three-axis fluxgate magnetometers with nonlinear least square method
journal, May 2013


Influence of field inhomogeneity in magnetic calibration coils
journal, January 1996

  • Nissen, J.; Paulsson, L. -E.
  • IEEE Transactions on Instrumentation and Measurement, Vol. 45, Issue 1
  • DOI: 10.1109/19.481353