Title: A prototype scintillator real‐time beam monitor for ultra‐high dose rate radiotherapy

Journal Article · · Medical Physics
DOI: https://doi.org/10.1002/mp.17018 · OSTI ID:2575911
 [1];  [2];  [1];  [1];  [1];  [1];  [3];  [3]
  1. Univ. of Michigan, Ann Arbor, MI (United States)
  2. Integrated Sensors LLC, Palm Beach Gardens, FL (United States)
  3. Loma Lima Univ., Loma Linda, CA (United States)

Background: FLASH Radiotherapy (RT) is an emergent cancer RT modality where an entire therapeutic dose is delivered at more than 1000 times higher dose rate than conventional RT. For clinical trials to be conducted safely, a precise and fast beam monitor that can generate out-of-tolerance beam interrupts is required. This paper describes the overall concept and provides results from a prototype ultra-fast, scintillator-based beam monitor for both proton and electron beam FLASH applications. Purpose: A FLASH Beam Scintillator Monitor (FBSM) is being developed that employs a novel proprietary scintillator material. The FBSM has capabilities that conventional RT detector technologies are unable to simultaneously provide: (1) large area coverage; (2) a low mass profile; (3) a linear response over a broad dynamic range; (4) radiation hardness; (5) real-time analysis to provide an IEC-compliant fast beam-interrupt signal based on true two-dimensional beam imaging, radiation dosimetry and excellent spatial resolution. Methods: The FBSM uses a proprietary low mass, less than 0.5 mm water equivalent, non-hygroscopic, radiation tolerant scintillator material (designated HM: hybrid material) that is viewed by high frame rate CMOS cameras. Folded optics using mirrors enable a thin monitor profile of ∼10 cm. A field programmable gate array (FPGA) data acquisition system generates real-time analysis on a time scale appropriate to the FLASH RT beam modality: 100–1000 Hz for pulsed electrons and 10–20 kHz for quasi-continuous scanning proton pencil beams. An ion beam monitor served as the initial development platform for this work and was tested in low energy heavy-ion beams (86Kr+26 and protons). A prototype FBSM was fabricated and then tested in various radiation beams that included FLASH level dose per pulse electron beams, and a hospital RT clinic with electron beams. Results: Results presented in this report include image quality, response linearity, radiation hardness, spatial resolution, and real-time data processing. Furthermore, the HM scintillator was found to be highly radiation damage resistant. It exhibited a small 0.025%/kGy signal decrease from a 216 kGy cumulative dose resulting from continuous exposure for 15 min at a FLASH compatible dose rate of 237 Gy/s. Measurements of the signal amplitude versus beam fluence demonstrate linear response of the FBSM at FLASH compatible dose rates of >40 Gy/s. Comparison with commercial Gafchromic film indicates that the FBSM produces a high resolution 2D beam image and can reproduce a nearly identical beam profile, including primary beam tails. The spatial resolution was measured at 35–40 µm. Tests of the firmware beta version show successful operation at 20 000 Hz frame rate or 50 µs/frame, where the real-time analysis of the beam parameters is achieved in less than 1 µs. Conclusions: The FBSM is designed to provide real-time beam profile monitoring over a large active area without significantly degrading the beam quality. A prototype device has been staged in particle beams at currents of single particles up to FLASH level dose rates, using both continuous ion beams and pulsed electron beams. Using a novel scintillator, beam profiling has been demonstrated for currents extending from single particles to 10 nA currents. Radiation damage is minimal and even under FLASH conditions would require ≥50 kGy of accumulated exposure in a single spot to result in a 1% decrease in signal output. Beam imaging is comparable to radiochromic films, and provides immediate images without hours of processing. Real-time data processing, taking less than 50 µs (combined data transfer and analysis times), has been implemented in firmware for 20 kHz frame rates for continuous proton beams.

Research Organization:
Integrated Sensors, LLC, Ottawa Hills, OH (United States)
Sponsoring Organization:
National Institutes of Health (NIH); USDOE Office of Science (SC), Nuclear Physics (NP)
Grant/Contract Number:
SC0019597
OSTI ID:
2575911
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 4 Vol. 51; ISSN 2473-4209; ISSN 0094-2405
Publisher:
American Association of Physicists in MedicineCopyright Statement
Country of Publication:
United States
Language:
English

References (34)

Proton FLASH Radiotherapy for the Treatment of Symptomatic Bone Metastases journal January 2023
Implementation and validation of a beam‐current transformer on a medical pulsed electron beam LINAC for FLASH‐RT beam monitoring journal October 2021
Dual beam‐current transformer design for monitoring and reporting of electron ultra‐high dose rate (FLASH) beam parameters journal January 2023
Ultra‐high dose rate electron beams and the FLASH effect: From preclinical evidence to a new radiotherapy paradigm journal January 2022
Technical note: Validation of an ultrahigh dose rate pulsed electron beam monitoring system using a current transformer for FLASH preclinical studies journal February 2022
A 2D strip ionization chamber array with high spatiotemporal resolution for proton pencil beam scanning FLASH radiotherapy journal May 2022
Synchronized high‐speed scintillation imaging of proton beams, generated by a gantry‐mounted synchrocyclotron, on a pulse‐by‐pulse basis journal July 2022
Time‐resolved dose rate measurements in pencil beam scanning proton FLASH therapy with a fiber‐coupled scintillator detector system journal December 2022
Geant4—a simulation toolkit
  • Agostinelli, S.; Allison, J.; Amako, K.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 506, Issue 3 https://doi.org/10.1016/S0168-9002(03)01368-8
journal July 2003
Radiation damage in scintillating crystals journal August 1998
Technical challenges for FLASH proton therapy journal October 2020
Perspectives in linear accelerator for FLASH VHEE: Study of a compact C-band system journal December 2022
Measuring the absolute light yield of scintillators
  • de Haas, J. T. M.; Dorenbos, P.; van Eijk, C. W. E.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 537, Issue 1-2 https://doi.org/10.1016/j.nima.2004.07.243
journal January 2005
1 kHz laser accelerated electron beam feasible for radiotherapy uses: A PIC–Monte Carlo based study
  • Polanek, R.; Hafz, Nasr A. M.; Lécz, Zs.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 987 https://doi.org/10.1016/j.nima.2020.164841
journal January 2021
SRIM – The stopping and range of ions in matter (2010)
  • Ziegler, James F.; Ziegler, M. D.; Biersack, J. P.
  • Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, Vol. 268, Issue 11-12 https://doi.org/10.1016/j.nimb.2010.02.091
journal June 2010
Pencil beam scanning proton FLASH maintains tumor control while normal tissue damage is reduced in a mouse model journal October 2022
Nanoporous aerogel-based periodic high-energy electron current x-ray sensors journal May 2020
Characterization of a new scintillation imaging system for proton pencil beam dose rate measurements journal August 2020
A multi-layer strip ionization chamber (MLSIC) device for proton pencil beam scan quality assurance journal August 2022
Ultra-fast, high spatial resolution single-pulse scintillation imaging of synchrocyclotron pencil beam scanning proton delivery journal February 2023
Development of Ultra-High Dose-Rate (FLASH) Particle Therapy journal March 2022
Maximum proton kinetic energy and patient‐generated neutron fluence considerations in proton beam arc delivery radiation therapy journal January 2009
An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan journal September 2012
AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams journal September 1999
Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice journal July 2014
Statistical Analysis of Fast Neutron Total Cross Sections of Silicon, Phosphor, Sulfur, and Chlorine journal May 1963
Hypofractionated FLASH-RT as an Effective Treatment against Glioblastoma that Reduces Neurocognitive Side Effects in Mice journal February 2021
Shining a FLASHlight on Ultrahigh Dose-Rate Radiation and Possible Late Toxicity journal June 2022
FLASH Radiotherapy: History and Future journal May 2021
Establishment and Initial Experience of Clinical FLASH Radiotherapy in Canine Cancer Patients journal May 2021
Peripheral Organ Equivalent Dose Estimation Procedure in Proton Therapy journal May 2022
FLASH Proton Pencil Beam Scanning Irradiation Minimizes Radiation-Induced Leg Contracture and Skin Toxicity in Mice journal March 2021
FLASH radiotherapy: an emerging approach in radiation therapy journal May 2022
Systematic Study of Inorganic and Organic Scintillator Light Yields conference November 2016