DOE PAGES title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: In vivo evaluation of an adaptive resuscitation controller using whole blood and crystalloid infusates for hemorrhagic shock

Journal Article · · Frontiers in Bioengineering and Biotechnology

Introduction Hemorrhage remains the leading cause of preventable death on the battlefield. The most effective means to increase survivability is early hemorrhage control and fluid resuscitation. Unfortunately, fluid resuscitation requires constant adjustments to ensure casualty is properly managed, which is often not feasible in the pre-hospital setting. In this study, we showed how an adaptive closed-loop controller for hemorrhage resuscitation can be used to automate hemodynamic management using a swine hemorrhagic shock injury model. Methods The adaptive resuscitation controller (ARC) was previously developed to track pressure–volume responsiveness in real time and adjust its infusion rate to reach the target mean arterial pressure (MAP). Swine while maintained under a surgical plane of anesthesia and analgesia underwent a splenectomy, followed by two hemorrhage and resuscitation events. For the first resuscitation event, hemorrhage was induced to reduce the MAP to 35 mmHg until arterial lactate reached 4 mmol/L. The ARC system then infused whole blood (WB) to reach the target MAP and maintained the subject using crystalloids for 120 min. For the second resuscitation event, the subjects were hemorrhaged again but resuscitated using only crystalloid infusion to reach the target MAP and 120-min maintenance. Results The ARC was effective at WB resuscitation, reaching the target MAP in 2.0 ± 1.0 min. The median performance error was 1.1% ± 4.6%, and target overshoot was 14.4% ± 7.0% of the target MAP. The ARC maintained all animals throughout the 120 min maintenance period. For the second crystalloid-based resuscitation, ARC required a longer time to reach the target MAP, at an average rise time of 4.3 ± 4.0 min. However, target overshoot was reduced to 8.4% ± 7.3% of the target MAP. Much higher flow rates were required to maintain the target MAP during the second resuscitation event than during the first resuscitation event. Discussion The ARC was able to rapidly reach and maintain the target MAP effectively. However, this sometimes required large volumes of fluid as the ARC’s only goal was to reach the target MAP. Further clinical insight is needed regarding the preferred aggression level to achieve the target MAP. In conclusion, the ARC was successful in its programmed objective of reaching and maintaining the target MAP for extended periods of time in vivo , a critical next step toward improving hemorrhage treatment in the pre-hospital environment.

Sponsoring Organization:
USDOE
OSTI ID:
2476717
Journal Information:
Frontiers in Bioengineering and Biotechnology, Journal Name: Frontiers in Bioengineering and Biotechnology Vol. 12; ISSN 2296-4185
Publisher:
Frontiers Media SACopyright Statement
Country of Publication:
Switzerland
Language:
English

References (35)

Robotics and Manufacturing Automation book December 2023
Measuring the predictive performance of computer-controlled infusion pumps journal February 1992
Performance of closed-loop resuscitation in a pig model of haemorrhagic shock with fluid alone or in combination with norepinephrine, a pilot study journal June 2020
Modern Missile Flight Control Design: An Overview journal September 2001
Monitoring fluid responsiveness journal June 2011
Data-driven based HVAC optimisation approaches: A Systematic Literature Review journal April 2022
Time Efficient Robust PID Plus Controller for Satellite Attitude Stabilization Control Considering Angular Velocity and Control Torque Constraint journal September 2017
Development of the PhysioVessel: a customizable platform for simulating physiological fluid resuscitation journal April 2022
Damage Control Resuscitation journal September 2018
Whole Blood Transfusion journal September 2018
Establishing a Regional Trauma Preventable/Potentially Preventable Death Rate journal February 2020
Closed-Loop Control of Fluid Therapy for Treatment of Hypovolemia journal April 2008
Death on the battlefield (2001–2011): Implications for the future of combat casualty care journal January 2012
Uncontrolled Hemorrhage Differs from Volume- or Pressure-Matched Controlled Hemorrhage in Swine journal October 2007
Hemodynamic Monitoring via Model-Based Extended Kalman Filtering: Hemorrhage Resuscitation and Sedation Case Study journal January 2022
Hemodynamic Safety Assurance in Closed-Loop Controlled Critical Care: Hemorrhage Resuscitation and Sedation Case Study journal January 2023
Evaluation of Fluid Resuscitation Control Algorithms via a Hardware-in-the-Loop Test Bed journal February 2020
A Population-Informed Particle Filter for Robust Physiological Monitoring Using Low-Information Time-Series Measurements journal August 2023
The effects of low‐dose epinephrine infusion on spleen size, central and hepatic circulation and circulating platelets journal August 2012
Adaptive closed‐loop resuscitation controllers for hemorrhagic shock resuscitation journal May 2023
Closed-Loop Fluid Resuscitation Control Via Blood Volume Estimation journal July 2016
Pathophysiology of Hemorrhage as It Relates to the Warfighter journal May 2022
Abstract 191: Effects of Anesthetic Agents on Blood Parameters in Rats With Acute Hemorrhagic Shock journal November 2018
Arterial Pulse Pressure Variation with Mechanical Ventilation journal January 2019
Different Effects of Volatile and Nonvolatile Anesthetic Agents on Long-Term Survival in an Experimental Model of Hemorrhagic Shock journal April 2020
Evaluation of a novel closed-loop fluid-administration system based on dynamic predictors of fluid responsiveness: an in silico simulation study journal November 2011
Closed-loop assisted versus manual goal-directed fluid therapy during high-risk abdominal surgery: a case–control study with propensity matching journal December 2015
Automated closed-loop resuscitation of multiple hemorrhages: a comparison between fuzzy logic and decision table controllers in a sheep model journal January 2017
Damage control resuscitation journal March 2020
Provisional Mortality Data — United States, 2022 journal May 2023
An Automated Hardware-in-Loop Testbed for Evaluating Hemorrhagic Shock Resuscitation Controllers journal August 2022
Hardware-in-Loop Comparison of Physiological Closed-Loop Controllers for the Autonomous Management of Hypotension journal August 2022
Closed-Loop Controlled Fluid Administration Systems: A Comprehensive Scoping Review journal July 2022
Implementation and Evaluation of Open-Source Hardware to Monitor Water Quality in Precision Aquaculture journal October 2020
Supervisory Algorithm for Autonomous Hemodynamic Management Systems journal January 2022