Title: On the perturbation effect and LET dependence of beam quality correction factors in carbon ion beams

Journal Article · · Medical Physics
DOI: https://doi.org/10.1002/mp.16089 · OSTI ID:2424833

Background: In a recent study, we reported beam quality correction factors, fQ, in carbon ion beams using Monte Carlo (MC) methods for a cylindrical and a parallel-plate ionization chamber (IC). A non-negligible perturbation effect was observed; however, the magnitude of the perturbation correction due to the specific IC subcomponents was not included. Furthermore, the stopping power data presented in the International Commission on Radiation Units and Measurements (ICRU) report 73 were used, whereas the latest stopping power data have been reported in the ICRU report 90. Purpose: The aim of this study was to extend our previous work by computing fQ correction factors using the ICRU 90 stopping power data and by reporting IC-specific perturbation correction factors. Possible energy or linear energy transfer (LET) dependence of the fQ correction factor was investigated by simulating both pristine beams and spread-out Bragg peaks (SOBPs). Methods: The TOol for PArticle Simulation (TOPAS)/GEANT4 MC code was used in this study. A 30 × 30 × 50 cm3 water phantom was simulated with a uniform 10 × 10 cm2 parallel beam incident on the surface. A Farmer-type cylindrical IC (Exradin A12) and two parallel-plate ICs (Exradin P11 and A11) were simulated in TOPAS using the manufacturer-provided geometrical drawings. The fQ correction factor was calculated in pristine carbon ion beams in the 150–450 MeV/u energy range at 2 cm depth and in the middle of the flat region of four SOBPs. The kQ correction factor was calculated by simulating the fQo correction factor in a 60Co beam at 5 cm depth. The perturbation correction factors due to the presence of the individual IC subcomponents, such as the displacement effect in the air cavity, collecting electrode, chamber wall, and chamber stem, were calculated at 2 cm depth for monoenergetic beams only. Additionally, the mean dose-averaged and track-averaged LET was calculated at the depths at which the fQ was calculated. Results: The ICRU 90 fQ correction factors were reported. The pdis correction factor was found to be significant for the cylindrical IC with magnitudes up to 1.70%. The individual perturbation corrections for the parallel-plate ICs were <1.0% except for the A11 pcel correction at the lowest energy. The fQ correction for the P11 IC exhibited an energy dependence of >1.00% and displayed differences up to 0.87% between pristine beams and SOBPs. Conversely, the fQ for A11 and A12 displayed a minimal energy dependence of <0.50%. The energy dependence was found to manifest in the LET dependence for the P11 IC. A statistically significant LET dependence was found only for the P11 IC in pristine beams only with a magnitude of <1.10%. Conclusions: The perturbation and kQ correction factor should be calculated for the specific IC to be used in carbon ion beam reference dosimetry as a function of beam quality.

Research Organization:
University of Wisconsin, Madison, WI (United States)
Sponsoring Organization:
National Science Foundation (NSF); USDOE Office of Science (SC)
OSTI ID:
2424833
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 2 Vol. 50; ISSN 0094-2405
Publisher:
American Association of Physicists in MedicineCopyright Statement
Country of Publication:
United States
Language:
English

References (20)

The influence of nuclear interactions on ionization chamber perturbation factors in proton beams: FLUKA simulations supported by a Fano test journal December 2018
Comparison of penh , fluka , and Geant4/ topas for absorbed dose calculations in air cavities representing ionization chambers in high‐energy photon and proton beams journal August 2019
Report on G4‐Med, a Geant4 benchmarking system for medical physics applications developed by the Geant4 Medical Simulation Benchmarking Group journal December 2020
Monte Carlo simulated beam quality and perturbation correction factors for ionization chambers in monoenergetic proton beams journal October 2020
Monte Carlo computation of 3D distributions of stopping power ratios in light ion beam therapy using GATE‐RTion journal March 2021
Geant4—a simulation toolkit journal July 2003
Validation of recent Geant4 physics models for application in carbon ion therapy
  • Lechner, A.; Ivanchenko, V. N.; Knobloch, J.
  • Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, Vol. 268, Issue 14 https://doi.org/10.1016/j.nimb.2010.04.008
journal July 2010
A systematic review on the usage of averaged LET in radiation biology for particle therapy journal August 2021
Monte Carlo calculation of perturbation correction factors for air-filled ionization chambers in clinical proton beams using TOPAS/GEANT journal May 2021
Bragg-Gray theory and ion chamber dosimetry for photon beams journal April 1991
Impact of new ICRU 90 key data on stopping-power ratios and beam quality correction factors for carbon ion beams journal September 2019
Report 49 journal May 1993
Geant4 developments and applications journal February 2006
Monte Carlo calculations of , the beam quality conversion factor journal October 2010
Beam quality conversion factors for parallel-plate ionization chambers in MV photon beams journal March 2012
TOPAS: An innovative proton Monte Carlo platform for research and clinical applications: TOPAS: An innovative proton Monte Carlo platform journal October 2012
Addendum to the AAPMˈs TG‐51 protocol for clinical reference dosimetry of high‐energy photon beams journal March 2014
AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams journal September 1999
Monte Carlo simulation of a typical 60Co therapy source journal November 1999
A Theory of Cavity Ionization journal November 1955