Title: Theoretical and experimental study on the detection limit of the micro-ring resonator based ultrasound point detectors

Journal Article · · Photoacoustics

Combining the diffusive laser excitation and the photoacoustic signals detection, photoacoustic computed tomography (PACT) is uniquely suited for deep tissue imaging. A diffraction-limited ultrasound point detector is highly desirable for maximizing the spatial resolution and the field-of-view of the reconstructed volumetric images. Among all the available ultrasound detectors, micro-ring resonator (MRR) based ultrasound detectors offer the lowest area-normalized limit of detection (nLOD) in a miniature form-factor, making it an ideal candidate as an ultrasound point detector. However, despite their wide adoption for photoacoustic imaging, the underlying signal transduction process has not been systematically studied yet. Here we report a comprehensive theoretical model capturing the transduction of incident acoustic signals into digital data, and the associated noise propagation process, using experimentally calibrated key process parameters. The theoretical model quantifies the signal-to-noise ratio (SNR) and the nLOD under the influence of the key process variables, including the quality factor (Q-factor) of the MRR and the driving wavelength. While asserting the need for higher Q-factors, the theoretical model further quantifies the optimal driving wavelength for optimizing the nLOD. Given the MRR with a Q-factor of 1 × 105, the theoretical model predicts an optimal SNR of 30.1 dB and a corresponding nLOD of 3.75 × 10-2 mPa mm2/Hz1/2, which are in good agreement with the experimental measurements of 31.0 dB and 3.39 × 10-2 mPa mm2/Hz1/2, respectively. The reported theoretical model can be used in guiding the optimization of MRR-based ultrasonic detectors and PA experimental conditions, in attaining higher imaging resolution and contrast. The optimized operating condition has been further validated by performing PACT imaging of a human hair phantom.

Research Organization:
Argonne National Laboratory (ANL), Argonne, IL (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Basic Energy Sciences (BES). Scientific User Facilities (SUF); National Institutes of Health (NIH); National Science Foundation (NSF); USDOE Office of Science (SC), Basic Energy Sciences (BES)
Grant/Contract Number:
AC02-06CH11357; P41GM135018; R01GM143397; R01EB028143; R01NS111039; RF1NS115581; R01HL166522; 2144788; 226178
OSTI ID:
2222832
Alternate ID(s):
OSTI ID: 2469530; OSTI ID: 2513857
Journal Information:
Photoacoustics, Vol. 34; ISSN 2213-5979
Publisher:
ElsevierCopyright Statement
Country of Publication:
United States
Language:
English

References (37)

Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries journal January 2008
A Review of Transparent Sensors for Photoacoustic Imaging Applications journal August 2021
Spatial resolution in photoacoustic tomography: effects of detector size and detector bandwidth journal October 2010
High-frequency ultrasound sensors using polymer microring resonators journal May 2007
Photoacoustic probe using a microring resonator ultrasonic sensor for endoscopic applications journal July 2014
Photoacoustic Tomography: In Vivo Imaging from Organelles to Organs journal March 2012
Photoacoustic computed tomography by using a multi-angle scanning fiber-laser ultrasound sensor journal January 2020
Real-time GPU-accelerated processing and volumetric display for wide-field laser-scanning optical-resolution photoacoustic microscopy journal January 2015
Highly sensitive ultrasound detection using nanofabricated polymer micro-ring resonators journal June 2023
Double-Stage Delay Multiply and Sum Beamforming Algorithm: Application to Linear-Array Photoacoustic Imaging journal January 2018
Deep in vivo photoacoustic imaging of mammalian tissues using a tyrosinase-based genetic reporter journal March 2015
Isometric multimodal photoacoustic microscopy based on optically transparent micro-ring ultrasonic detection journal January 2015
Quadruple ultrasound, photoacoustic, optical coherence, and fluorescence fusion imaging with a transparent ultrasound transducer journal March 2021
Multiscale photoacoustic microscopy and computed tomography journal August 2009
Lithium niobate-based transparent ultrasound transducers for photoacoustic imaging journal October 2019
Fabrication and characterization of High Q polymer micro-ring resonator and its application as a sensitive ultrasonic detector journal January 2011
High-sensitivity and wide-directivity ultrasound detection using high Q polymer microring resonators journal May 2011
Non-contact biomedical photoacoustic and ultrasound imaging journal January 2012
Ultrasensitive Non-Resonant Detection of Ultrasound with Plasmonic Metamaterials journal March 2013
Ultrasonic detection based on polarization-dependent optical reflection journal January 2017
Broadband optical ultrasound sensor with a unique open-cavity structure journal January 2011
Silicon microring resonators journal September 2011
A submicrometre silicon-on-insulator resonator for ultrasound detection journal September 2020
Looking at sound: optoacoustics with all-optical ultrasound detection journal August 2018
Optical Detection of Ultrasound in Photoacoustic Imaging journal January 2017
Advances in real-time multispectral optoacoustic imaging and its applications journal March 2015
In vivo three-dimensional photoacoustic tomography of a whole mouse head journal July 2006
Rapid volumetric photoacoustic tomographic imaging with a Fabry-Perot ultrasound sensor depicts peripheral arteries and microvascular vasomotor responses to thermal stimuli journal October 2017
Ultrasensitive plano-concave optical microresonators for ultrasound sensing journal October 2017
High-sensitivity compact ultrasonic detector based on a pi-phase-shifted fiber Bragg grating journal May 2011
Performance and noise analysis of optical microresonator-based biochemical sensors using intensity detection journal July 2016
A photoacoustic finder fully integrated with a solid-state dye laser and transparent ultrasound transducer journal September 2021
Photoacoustic Microscopy and Computed Tomography: From Bench to Bedside journal July 2014
High-Frequency 3D Photoacoustic Computed Tomography Using an Optical Microring Resonator journal January 2022
PHOTOACOUSTIC TOMOGRAPHY: PRINCIPLES AND ADVANCES (Invited Review) journal January 2014
Silicon‐Photonics Point Sensor for High‐Resolution Optoacoustic Imaging journal July 2021
Backward-mode multiwavelength photoacoustic scanner using a planar Fabry-Perot polymer film ultrasound sensor for high-resolution three-dimensional imaging of biological tissues journal January 2008