Title: Computational models of direct and indirect X‐ray breast imaging detectors for in silico trials

Journal Article · · Medical Physics
DOI: https://doi.org/10.1002/mp.15935 · OSTI ID:1902568
 [1];  [1];  [1];  [1]
  1. Division of Imaging Diagnostics and Software Reliability Office of Science and Engineering Laboratories Center for Devices and Radiological Health U. S. Food and Drug Administration Silver Spring Maryland USA

Abstract Background To facilitate in silico studies that investigate digital mammography (DM) and breast tomosynthesis (DBT), models replicating the variety in imaging performance of the DM and DBT systems, observed across manufacturers are needed. Purpose The main purpose of this work is to develop generic physics models for direct and indirect detector technology used in commercially available systems, with the goal of making them available open source to manufacturers to further tweak and develop the exact in silico replicas of their systems. Methods We recently reported on an in silico version of the SIEMENS Mammomat Inspiration DM/DBT system using an open‐source GPU‐accelerated Monte Carlo x‐ray imaging simulation code (MC‐GPU). We build on the previous version of the MC‐GPU codes to mimic the imaging performances of two other Food and Drug Administration (FDA)‐approved DM/DBT systems, such as Hologic Selenia Dimensions (HSD) and the General Electric Senographe Pristina (GSP) systems. In this work, we developed a hybrid technique to model the optical spread and signal crosstalk observed in the GSP and HSD systems. MC simulations are used to track each x‐ray photon till its first interaction within the x‐ray detector. On the other hand, the signal spread in the x‐ray detectors is modeled using previously developed analytical equations. This approach allows us to preserve the modeling accuracy offered by MC methods in the patient body, while speeding up secondary carrier transport (either electron–hole pairs or optical photons) using analytical equations in the detector. The analytical optical spread model for the indirect detector includes the depth‐dependent spread and collection of optical photons and relies on a pre‐computed set of point response functions that describe the optical spread as a function of depth. To understand the capabilities of the computational x‐ray detector models, we compared image quality metrics like modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE), simulated with our models against measured data. Please note that the purpose of these comparisons with measured data would be to gauge if the model developed as part of this work could replicate commercially used direct and indirect technology in general and not to achieve perfect fits with measured data. Results We found that the simulated image quality metrics such as MTF, NNPS, and DQE were in reasonable agreement with experimental data. To demonstrate the imaging performance of the three DM/DBT systems, we integrated the detector models with the VICTRE pipeline and simulated DM images of a fatty breast model containing a spiculated mass and a calcium oxalate cluster. In general, we found that the images generated using the indirect model appeared more blurred with a different noise texture and contrast as compared to the systems with direct detectors. Conclusions We have presented computational models of three commercially available FDA‐approved DM/DBT systems, which implement both direct and indirect detector technology. The updated versions of the MC‐GPU codes that can be used to replicate three systems are available in open source format through GitHub.

Sponsoring Organization:
USDOE
OSTI ID:
1902568
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 11 Vol. 49; ISSN 0094-2405
Publisher:
Wiley Blackwell (John Wiley & Sons)Copyright Statement
Country of Publication:
United States
Language:
English

References (22)

Evaluation of Digital Breast Tomosynthesis as Replacement of Full-Field Digital Mammography Using an In Silico Imaging Trial journal November 2018
Comparing sensitivity and specificity of screening mammography in the United States and Denmark journal June 2015
Deriving depth-dependent light escape efficiency and optical Swank factor from measured pulse height spectra of scintillators journal February 2017
Technical Note: In silico imaging tools from the VICTRE clinical trial journal July 2019
Breast cancer detection rates using four different types of mammography detectors journal June 2015
Quantitative Image Quality Measurements of a Digital Breast Tomosynthesis System journal July 2013
MANTIS: combined x-ray, electron and optical Monte Carlo simulations of indirect radiation imaging systems journal March 2006
hybrid$\scriptsize{\mathrm{MANTIS}}$: a CPU–GPU Monte Carlo method for modeling indirect x-ray detectors with columnar scintillators journal April 2012
Characterisation of noise and sharpness of images from four digital breast tomosynthesis systems for simulation of images for virtual clinical trials journal February 2017
Cascaded systems analysis of a-Se/a-Si and a-InGaZnO TFT passive and active pixel sensors for tomosynthesis journal January 2019
DynAMITe: a wafer scale sensor for biomedical applications journal December 2011
Cascade Modeling of Pixelated Scintillator Detectors for X-Ray Imaging journal June 2008
A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems journal May 2010
Oblique incidence effects in direct x‐ray detectors: A first‐order approximation using a physics‐based analytical model journal March 2011
Spatiotemporal Monte Carlo transport methods in x-ray semiconductor detectors: Application to pulse-height spectroscopy in a-Se journal December 2011
A review of breast tomosynthesis. Part I. The image acquisition process: Breast tomosynthesis review. I. Image acquisition journal January 2013
Validation of columnar CsI x-ray detector responses obtained with hybrid MANTIS , a CPU-GPU Monte Carlo code for coupled x-ray, electron, and optical transport : Validation of hybrid journal February 2013
Free software for performing physical analysis of systems for digital radiography and mammography journal April 2014
Large area CMOS active pixel sensor x‐ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization journal October 2015
X-ray imaging using amorphous selenium: Inherent spatial resolution journal April 1995
A method for measuring the presampled MTF of digital radiographic systems using an edge test device journal January 1998
Individual and Combined Effects of Age, Breast Density, and Hormone Replacement Therapy Use on the Accuracy of Screening Mammography journal February 2003