Title: Absolute calibration of the conical crystal configuration of the zinc spectrometer (ZSPEC) at the OMEGA laser facility

Journal Article · · Review of Scientific Instruments
DOI: https://doi.org/10.1063/5.0099176 · OSTI ID:1889251

In this study, we present the absolute calibration of the conical crystal for the zinc spectrometer (ZSPEC), an x-ray spectrometer at the OMEGA laser facility at the Laboratory for Laser Energetics. The ZSPEC was originally designed to measure x-ray Thomson scattering using flat or cylindrically curved highly oriented pyrolytic graphite crystals centered around Zn He-alpha emission at 9 keV. To improve the useful spectral range and collection efficiency of the ZSPEC, a conical highly annealed pyrolytic graphite crystal was fabricated for the ZSPEC. The conically bent crystal in the Hall geometry produces a line focus perpendicular to the spectrometer axis, corresponding to the detector plane of electronic detectors at large scale laser facilities such as OMEGA, extending the useful range of the spectrometer to 7–11 keV. Further, using data collected using a microfocus Mo x-ray source, we determine important characteristics of ZSPEC such as the dispersion, spatial resolution, and absolute sensitivity of the instrument. A ray-trace model of ZSPEC provides another point of agreement in calculations of the ZSPEC dispersion and crystal response.

Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States); Princeton Plasma Physics Laboratory (PPPL), Princeton, NJ (United States)
Sponsoring Organization:
USDOE; USDOE Laboratory Directed Research and Development (LDRD) Program; USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
AC52-07NA27344
OSTI ID:
1889251
Report Number(s):
LLNL-JRNL-835280; 18-ERD-033; 22-ERD-005
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 8 Vol. 93; ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)Copyright Statement
Country of Publication:
United States
Language:
English

References (10)

Smoothing and Differentiation of Data by Simplified Least Squares Procedures. journal July 1964
System for calibrating the energy-dependent response of an elliptical Bragg-crystal spectrometer journal November 2014
X-ray Thomson scattering measurements from hohlraum-driven spheres on the OMEGA laser journal August 2016
An x-ray optic calibration facility for high energy density diagnostics journal October 2018
Developing a long-duration Zn K-α source for x-ray scattering experiments journal October 2018
Design and expected performance of a variable-radii sinusoidal spiral x-ray spectrometer for the National Ignition Facility journal September 2021
A focusing X-ray crystal spectrograph journal February 1984
Absolute calibration of the continuum x-ray spectrometer (ConSpec) at the National Ignition Facility journal December 2019
On the design of the NIF Continuum Spectrometer conference January 2018
XOP v2.4: recent developments of the x-ray optics software toolkit conference September 2011