DOE PAGES title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A Monte Carlo technique to model performance of streak camera-based time-resolving x-ray spectrometers

Journal Article · · Review of Scientific Instruments
DOI: https://doi.org/10.1063/5.0101705 · OSTI ID:1888051

A Monte Carlo technique has been developed to simulate the expected signal and the statistical noise of x-ray spectrometers that use streak cameras to achieve the time resolution required for ultrafast diagnostics of laser-generated plasmas. The technique accounts for statistics from both the photons incident on the streak camera’s photocathode and the electrons emitted by the photocathode travelling through the camera’s electron optics to the sensor. We use the technique to optimize the design of a spectrometer, which deduces the temporal history of electron temperature of the hotspot in an inertial confinement fusion implosion from its hard x-ray continuum emission spectra. The technique is general enough to be applied to any instrument using an x-ray streak camera.

Sponsoring Organization:
USDOE
Grant/Contract Number:
AC52-07NA27344
OSTI ID:
1888051
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 9 Vol. 93; ISSN 0034-6748
Publisher:
American Institute of PhysicsCopyright Statement
Country of Publication:
United States
Language:
English

References (9)

Ultrafast x‐ray streak camera for use in ultrashort laser‐produced plasma research journal January 1995
Performance of a CsI photocathode in a hard x-ray streak camera journal January 2000
The characterization of x‐ray photocathodes in the 0.1–10‐keV photon energy region journal March 1981
Improving the off-axis spatial resolution and dynamic range of the NIF X-ray streak cameras (invited) journal August 2016
Implementing time resolved electron temperature capability at the NIF using a streak camera journal October 2018
Electron emission from alkali halides under soft x-ray bombardment journal January 1973
Absolute calibration of the continuum x-ray spectrometer (ConSpec) at the National Ignition Facility journal December 2019
On the design of the NIF Continuum Spectrometer conference January 2018
Measuring x-ray burn history with the Streaked Polar Instrumentation for Diagnosing Energetic Radiation (SPIDER) at the National Ignition Facility (NIF) conference October 2012

Similar Records

Related Subjects