Title: Magnetic, Mechanical and Thermal Modeling of Superconducting, Whole-Body, Actively Shielded, 3 T MRI Magnets Wound Using MgB2 Strands for Liquid Cryogen Free Operation

Journal Article · · IEEE Transactions on Applied Superconductivity

Here we present magnetic, mechanical and thermal modeling results for a 3 Tesla actively shielded whole body MRI (Magnetic Resonance Imaging) magnet consisting of coils with a square cross section of their windings. The magnet design was a segmented coil type optimized to minimize conductor length while hitting the standard field quality and DSV (Diameter of Spherical Volume) specifications as well as a standard, compact size 3 T system. It had an overall magnet length and conductor length which can lead to conduction cooled designs comparable to NbTi helium bath cooled 3 T MRI magnets. The design had a magnetic field homogeneity better than 10 ppm (part-per-million) within a DSV (Diameter of Spherical Volume) of 48 cm and the total magnet winding length of 1.37 m. A new class of MgB2 strand especially designed for MRI applications was considered as a possible candidate for winding such magnets. This work represents the first magnetic, mechanical and thermal design for a whole-body 3 T MgB2 short (1.37 m length) MRI magnet based on the performance parameters of existing MgB2 wire. 3 Tesla MRI magnet can operate at 20 K at 67% of its critical current.

Research Organization:
Fermi National Accelerator Laboratory (FNAL), Batavia, IL (United States)
Sponsoring Organization:
National Institute of Biomedical Imaging and Bioengineering; National Institutes of Health (NIH); USDOE Office of Science (SC), High Energy Physics (HEP)
Grant/Contract Number:
AC02-07CH11359
OSTI ID:
1872455
Report Number(s):
FERMILAB-PUB-22-257-TD; oai:inspirehep.net:2064668
Journal Information:
IEEE Transactions on Applied Superconductivity, Journal Name: IEEE Transactions on Applied Superconductivity Journal Issue: 4 Vol. 32; ISSN 1051-8223
Publisher:
IEEECopyright Statement
Country of Publication:
United States
Language:
English

References (10)

Novel technologies and configurations of superconducting magnets for MRI journal July 2013
Conduction cooled magnet design for 1.5 T, 3.0 T and 7.0 T MRI systems journal November 2014
A multiscale and multiphysics model of strain development in a 1.5 T MRI magnet designed with 36 filament composite MgB2superconducting wire journal March 2016
Conductors for commercial MRI magnets beyond NbTi: requirements and challenges journal November 2016
Conceptual designs of conduction cooled MgB2 magnets for 1.5 and 3.0 T full body MRI systems journal March 2017
Instrumentation, cooling, and initial testing of a large, conduction-cooled, react-and-wind MgB2 coil segment for MRI applications journal July 2018
Quench, normal zone propagation velocity, and the development of an active protection scheme for a conduction cooled, react-and-wind, MgB2 MRI coil segment journal October 2019
High performance, advanced-internal-magnesium-infiltration (AIMI) MgB2 wires processed using a vapor-solid reaction route journal July 2020
Advances in Whole-Body MRI Magnets journal June 2011
Effect of Cold Work on the Thermal Conductivity of Copper journal September 1965

Similar Records

Construction and test results of a compact 0. 8 meter warm bore 1. 5 Tesla high homogeneity superconducting magnet for MR spectroscopy
Conference · 1991 · IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5257938

A 0. 6 Tesla superconducting magnet for magnetic resonance imaging
Conference · 1991 · IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5294763

Formation of homogeneous field in a shielded superconducting solenoid for an MRI
Conference · 1991 · IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5294404