DOE PAGES title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: The Rainwater Memorial Calibration Facility for X-Ray Optics

Journal Article · · X-Ray Optics and Instrumentation

The Nuclear Spectroscopic Telescope ARray (NuSTAR) is a NASA Small Explorer mission that will carry the first focusing hard X-ray (5–80 keV) telescope to orbit. The ground calibration of the optics posed a challenge as the need to suppress finite source distance effects over the full optic and the energy range of interest were unique requirements not met by any existing facility. In this paper we present the requirements for the NuSTAR optics ground calibration, and how the Rainwater Memorial Calibration Facility, RaMCaF, is designed to meet the calibration requirements. The nearly 175 m long beamline sports a 48 cm diameter 5–100 keV X-ray beam and is capable of carrying out detailed studies of large diameter optic elements, such as the NuSTAR optics, as well as flat multilayer-coated Silicon wafers.

Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States)
Sponsoring Organization:
NASA; USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
AC52-07NA27344
OSTI ID:
1865559
Report Number(s):
LLNL-JRNL-490180; 501551
Journal Information:
X-Ray Optics and Instrumentation, Journal Name: X-Ray Optics and Instrumentation Vol. 2011; ISSN 1687-7632
Publisher:
HindawiCopyright Statement
Country of Publication:
United States
Language:
English

References (3)

Construction and commissioning of a 215-m-long beamline at SPring-8 journal July 2001
Expanding the detection efficiency of silicon drift detectors
  • Schlosser, D. M.; Lechner, P.; Lutz, G.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 624, Issue 2 https://doi.org/10.1016/j.nima.2010.04.038
journal December 2010
IMD—Software for modeling the optical properties of multilayer films journal January 1998

Cited By (2)

Direct intensity calibration of X-ray grazing-incidence microscopes with home-lab source journal January 2018
An x-ray optic calibration facility for high energy density diagnostics journal October 2018