A custom high-stability pulse generator for the test of a novel in-situ magnetic sensor developed to detect radiation damage in undulators
Abstract
High-performance analog-to-digital converters (ADC) have been widely applied in many areas of science. For instance, magnetic field measurements based on the Faraday’s induction law require high-precision voltmeters to measure induced voltages. In this study in the context of free-electron lasers technology, the Magnetic Measurements Team at SLAC National Accelerator Laboratory proposed a novel in-situ radiation damage detection system (RDDS) for detecting small field variations in undulators. The system measures the flux change in a flexible printed-circuit coil attached to the magnet array during the undulator gap movement. The gap movement changes the magnetic field, which induces a voltage signal in the coil that is measured and integrated with an ADC. Although the system is capable of detecting relative flux changes better than 100 ppm, drift in the ADC’s gain or offset can cause apparent changes in the relative flux. This paper describes the first attempt to developed a high-precision verification circuit to perform ADC testing in the framework of the novel RDDS. The circuit generates a reference voltage pulse with a voltage–time integral relative precision better than 50 ppm for a few hundreds of mVs — the typical order of magnitude measured with the RDDS. The circuit’s design combines amore »
- Authors:
-
- European X-ray Free-Electron Laser (XFEL), Hamburg (Germany); SLAC National Accelerator Lab., Menlo Park, CA (United States)
- SLAC National Accelerator Lab., Menlo Park, CA (United States)
- Publication Date:
- Research Org.:
- SLAC National Accelerator Laboratory (SLAC), Menlo Park, CA (United States)
- Sponsoring Org.:
- USDOE Office of Science (SC), Basic Energy Sciences (BES); Coordenacao de Aperfeicoamento de Pessoal de Nivel Superior - Brazil (CAPES)
- OSTI Identifier:
- 1674944
- Alternate Identifier(s):
- OSTI ID: 1668125
- Grant/Contract Number:
- AC02-76SF00515
- Resource Type:
- Accepted Manuscript
- Journal Name:
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Additional Journal Information:
- Journal Volume: 985; Journal ID: ISSN 0168-9002
- Publisher:
- Elsevier
- Country of Publication:
- United States
- Language:
- English
- Subject:
- 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; analog-to-digital converters (ADC); ADC testing; magnetic field measurements; pulse generator; undulators
Citation Formats
Baader, Johann E., and Anderson, Scott D. A custom high-stability pulse generator for the test of a novel in-situ magnetic sensor developed to detect radiation damage in undulators. United States: N. p., 2021.
Web. doi:10.1016/j.nima.2020.164633.
Baader, Johann E., & Anderson, Scott D. A custom high-stability pulse generator for the test of a novel in-situ magnetic sensor developed to detect radiation damage in undulators. United States. https://doi.org/10.1016/j.nima.2020.164633
Baader, Johann E., and Anderson, Scott D. Fri .
"A custom high-stability pulse generator for the test of a novel in-situ magnetic sensor developed to detect radiation damage in undulators". United States. https://doi.org/10.1016/j.nima.2020.164633. https://www.osti.gov/servlets/purl/1674944.
@article{osti_1674944,
title = {A custom high-stability pulse generator for the test of a novel in-situ magnetic sensor developed to detect radiation damage in undulators},
author = {Baader, Johann E. and Anderson, Scott D.},
abstractNote = {High-performance analog-to-digital converters (ADC) have been widely applied in many areas of science. For instance, magnetic field measurements based on the Faraday’s induction law require high-precision voltmeters to measure induced voltages. In this study in the context of free-electron lasers technology, the Magnetic Measurements Team at SLAC National Accelerator Laboratory proposed a novel in-situ radiation damage detection system (RDDS) for detecting small field variations in undulators. The system measures the flux change in a flexible printed-circuit coil attached to the magnet array during the undulator gap movement. The gap movement changes the magnetic field, which induces a voltage signal in the coil that is measured and integrated with an ADC. Although the system is capable of detecting relative flux changes better than 100 ppm, drift in the ADC’s gain or offset can cause apparent changes in the relative flux. This paper describes the first attempt to developed a high-precision verification circuit to perform ADC testing in the framework of the novel RDDS. The circuit generates a reference voltage pulse with a voltage–time integral relative precision better than 50 ppm for a few hundreds of mVs — the typical order of magnitude measured with the RDDS. The circuit’s design combines a fast and precise switch with a low-noise voltage reference. Long-term measurements allowed statistical analysis and showed that averaging the voltage–time integral of ten pulses gives the required 50 ppm stability. Moreover, reproducibility tests confirmed that the circuit’s output is invariable under small power supply instabilities and equipment shutdown. Instruments and applications designed to quantify the magnetic field by integrating voltage signals may use the pulse generator proposed in this paper for verification purposes.},
doi = {10.1016/j.nima.2020.164633},
journal = {Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment},
number = ,
volume = 985,
place = {United States},
year = {Fri Jan 01 00:00:00 EST 2021},
month = {Fri Jan 01 00:00:00 EST 2021}
}
Works referenced in this record:
Custom pulse generator for RPC testing
journal, May 2009
- Gil, A.; Castro, E.; Díaz, J.
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 602, Issue 3
Panorama of new generation of accelerator based short wavelength coherent light sources
journal, December 2015
- Couprie, M. E.
- Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, Vol. 364
Reference pulsers for calibration purposes
journal, June 1974
- Francesco Manfredi, Pier; Maranesi, Piero; Varoli, Vincenzo
- Nuclear Instruments and Methods, Vol. 118, Issue 2
Search for possible radiation damage on a NdFeB permanent magnet structure after two years of operation
journal, February 1995
- Pflüger, J.; Heintze, G.; Vasserman, I.
- Review of Scientific Instruments, Vol. 66, Issue 2
Refurbishment of radiation-damaged undulators
conference, January 2016
- Tischer, M.; Neumann, P.; Schöps, A.
- PROCEEDINGS OF THE 12TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION – SRI2015, AIP Conference Proceedings
Random pulse generator with a uniformly distributed amplitude spectrum
journal, April 1988
- Lauch, J.; Nachbar, H. U.
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 267, Issue 1
Low-cost programmable pulse generator for particle telescope calibration
journal, November 1999
- Sánchez, S.; González, M.; Seisdedos, M.
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 436, Issue 3
A Novel $In-Situ$ Radiation Damage Diagnostic System for Undulators
journal, June 2020
- Baader, Johann E.; Anderson, Scott D.; Wolf, Zachary R.
- IEEE Transactions on Instrumentation and Measurement, Vol. 69, Issue 6
Precise calibration of a Ge(Li)-spectrometer using a digital to analog converter
journal, April 1971
- McKee, R. J.; Hargrove, C. K.; Smith, A. G.
- Nuclear Instruments and Methods, Vol. 92, Issue 3
Radiation-induced magnetization reversal causing a large flux loss in undulator permanent magnets
journal, November 2016
- Bizen, Teruhiko; Kinjo, Ryota; Hasegawa, Teruaki
- Scientific Reports, Vol. 6, Issue 1
Scott's rule: Scott's rule
journal, June 2010
- Scott, David W.
- Wiley Interdisciplinary Reviews: Computational Statistics, Vol. 2, Issue 4
A 4V, ns-range pulse generator for the test of Cherenkov Telescopes readout electronics
journal, August 2010
- Antoranz, P.; Vegas, I.; Miranda, J. M.
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 620, Issue 2-3