Performance of a Nonelectric Infant Warmer in Rwandan Health Centers
- Boston Children’s Hospital, MA (United States); Children's Hospital of Colorado, Aurora, CO (United States)
- Inshuti Mu Buzima, Kigali (Rwanda)
- Yale Univ., New Haven, CT (United States)
- Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
- Boston Children’s Hospital, MA (United States)
- Boston Children’s Hospital, MA (United States); Harvard Medical School, Boston, MA (United States)
Background: Neonatal hypothermia remains a challenge in resource-limited settings. Methods. We conducted a prospective mixed-methods cohort study in rural Rwandan health centers to assess the performance of an infant warmer we designed for low-resource settings. All hypothermic infants were eligible for enrollment. Outcomes. Safety: incidence of adverse reactions. Effectiveness: attainment of euthermia, rate of temperature rise. Feasibility: correct use of warmer, signs of wear. Interviews of caregivers and nurses. Findings. Of 102 encounters, there were no adverse reactions. Of 80 encounters for hypothermia when infants on warmer for ≥1 hour, 79 achieved euthermia; 73 in ≤2 hours. Of the 80 encounters, 64 had temperature rise ≥0.5°C/h. Of the 102 encounters, there were no instances of the warmer being prepared, used, or cleaned incorrectly. Five out of the 12 warmers exhibited wear. Interview participants were predominantly positive; some found time for readiness of warmer challenging. Interpretation. The warmer performed well. It is appropriate to study in larger scale.
- Research Organization:
- Lawrence Berkeley National Laboratory (LBNL), Berkeley, CA (United States)
- Sponsoring Organization:
- USDOE Office of Energy Efficiency and Renewable Energy (EERE); Harvard Medical School
- Grant/Contract Number:
- AC02-05CH11231
- OSTI ID:
- 1619112
- Journal Information:
- Global Pediatric Health, Journal Name: Global Pediatric Health Vol. 6; ISSN 2333-794X
- Publisher:
- SAGECopyright Statement
- Country of Publication:
- United States
- Language:
- English
Similar Records
A low cost, re-usable electricity-free infant warmer: evaluation of safety, effectiveness and feasibility
Physiologic implications of two different heat shields for infants under radiant warmers