A simple instrument to find spatiotemporal overlap of optical/X-ray light at free-electron lasers
Journal Article
·
· Journal of Synchrotron Radiation (Online)
- SLAC National Accelerator Lab., Menlo Park, CA (United States)
A compact and robust diagnostic to determine spatial and temporal overlap between X-ray free-electron laser and optical laser pulses was developed and evaluated using monochromatic X-rays from the Linac Coherent Light Source. It was used to determine temporal overlap with a resolution of ~10 fs, despite the large pulse energy fluctuations of the monochromatic X-ray pulses, and covers a wide optical wavelength range from ultraviolet to near-infrared with a single configuration.
- Research Organization:
- SLAC National Accelerator Laboratory (SLAC), Menlo Park, CA (United States)
- Sponsoring Organization:
- USDOE
- Grant/Contract Number:
- AC02-76SF00515
- OSTI ID:
- 1617936
- Alternate ID(s):
- OSTI ID: 1528882
- Journal Information:
- Journal of Synchrotron Radiation (Online), Vol. 26, Issue 3; ISSN 1600-5775
- Publisher:
- International Union of CrystallographyCopyright Statement
- Country of Publication:
- United States
- Language:
- English
Similar Records
An Optical Streaking Method for Measuring Femtosecond Electron Bunches
Pump–probe experimental methodology at the Linac Coherent Light Source
2D monochromatic x-ray imaging for beam monitoring of an x-ray free electron laser and a high-power femtosecond laser
Conference
·
2011
·
OSTI ID:1032758
Pump–probe experimental methodology at the Linac Coherent Light Source
Journal Article
·
2019
· Journal of Synchrotron Radiation (Online)
·
OSTI ID:1617931
+3 more
2D monochromatic x-ray imaging for beam monitoring of an x-ray free electron laser and a high-power femtosecond laser
Journal Article
·
2021
· Review of Scientific Instruments
·
OSTI ID:1630312
+16 more