skip to main content
DOE PAGES title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: On the system stability and calibration of the image plate/scanner system for plasma diagnosis at the National Ignition Facility

Abstract

At the National Ignition Facility (NIF), storage phosphor image plates (IP) are used extensively for recording x-rays, charged particles, and neutrons. For x-ray imaging and spectroscopy, absolute and relative calibrations are important for extracting plasma information from the diagnostics. We use Fuji MS, SR, and TR image plates that have been cut to fit custom diagnostic envelopes. The image plates are scanned on a General Electric FLA 7000 IP flying spot scanner. Calibrations for sensitivity, spatial scale, and temperature dependent fade are applied. During a set of recent calibrations, we noticed large shifts in the absolute calibration of the image plate system. The possible source of these shifts is discussed. We discuss scanner stability and a method for calibration. We discuss the fade and temperature effects of the image plates and how this correction is applied within the NIF environment. Here, we also compare our NIF GE FLA 7000 IP scanner with a new General Electric Amersham Typhoon IP scanner.

Authors:
 [1]; ORCiD logo [1];  [1];  [1];  [1]; ORCiD logo [1];  [1];  [1];  [1];  [1];  [1];  [1]
  1. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Publication Date:
Research Org.:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Org.:
USDOE National Nuclear Security Administration (NNSA)
OSTI Identifier:
1479074
Alternate Identifier(s):
OSTI ID: 1477907
Report Number(s):
LLNL-JRNL-750846
Journal ID: ISSN 0034-6748; 936509
Grant/Contract Number:  
AC52-07NA27344
Resource Type:
Accepted Manuscript
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 89; Journal Issue: 10; Journal ID: ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Country of Publication:
United States
Language:
English
Subject:
Lasers; 47 OTHER INSTRUMENTATION

Citation Formats

Holder, J. P., Izumi, N., Beach, M., Ayers, M. J., Bell, P., Schneider, M., Bradley, D. K., Kohut, T., Ehrlich, R., Cohen, M., Ramirez, R., and Thorn, D. On the system stability and calibration of the image plate/scanner system for plasma diagnosis at the National Ignition Facility. United States: N. p., 2018. Web. doi:10.1063/1.5039363.
Holder, J. P., Izumi, N., Beach, M., Ayers, M. J., Bell, P., Schneider, M., Bradley, D. K., Kohut, T., Ehrlich, R., Cohen, M., Ramirez, R., & Thorn, D. On the system stability and calibration of the image plate/scanner system for plasma diagnosis at the National Ignition Facility. United States. doi:https://doi.org/10.1063/1.5039363
Holder, J. P., Izumi, N., Beach, M., Ayers, M. J., Bell, P., Schneider, M., Bradley, D. K., Kohut, T., Ehrlich, R., Cohen, M., Ramirez, R., and Thorn, D. Wed . "On the system stability and calibration of the image plate/scanner system for plasma diagnosis at the National Ignition Facility". United States. doi:https://doi.org/10.1063/1.5039363. https://www.osti.gov/servlets/purl/1479074.
@article{osti_1479074,
title = {On the system stability and calibration of the image plate/scanner system for plasma diagnosis at the National Ignition Facility},
author = {Holder, J. P. and Izumi, N. and Beach, M. and Ayers, M. J. and Bell, P. and Schneider, M. and Bradley, D. K. and Kohut, T. and Ehrlich, R. and Cohen, M. and Ramirez, R. and Thorn, D.},
abstractNote = {At the National Ignition Facility (NIF), storage phosphor image plates (IP) are used extensively for recording x-rays, charged particles, and neutrons. For x-ray imaging and spectroscopy, absolute and relative calibrations are important for extracting plasma information from the diagnostics. We use Fuji MS, SR, and TR image plates that have been cut to fit custom diagnostic envelopes. The image plates are scanned on a General Electric FLA 7000 IP flying spot scanner. Calibrations for sensitivity, spatial scale, and temperature dependent fade are applied. During a set of recent calibrations, we noticed large shifts in the absolute calibration of the image plate system. The possible source of these shifts is discussed. We discuss scanner stability and a method for calibration. We discuss the fade and temperature effects of the image plates and how this correction is applied within the NIF environment. Here, we also compare our NIF GE FLA 7000 IP scanner with a new General Electric Amersham Typhoon IP scanner.},
doi = {10.1063/1.5039363},
journal = {Review of Scientific Instruments},
number = 10,
volume = 89,
place = {United States},
year = {2018},
month = {10}
}

Journal Article:
Free Publicly Available Full Text
Publisher's Version of Record

Citation Metrics:
Cited by: 2 works
Citation information provided by
Web of Science

Figures / Tables:

FIG. 1 FIG. 1: Schematic Energy Level Diagram of Typical Br F Ba Image Plate with Eu dopant/activator

Save / Share:

Works referenced in this record:

The physics of computed radiography
journal, November 2002


Application of imaging plates to x-ray imaging and spectroscopy in laser plasma experiments (invited)
journal, October 2006

  • Izumi, N.; Snavely, R.; Gregori, G.
  • Review of Scientific Instruments, Vol. 77, Issue 10
  • DOI: 10.1063/1.2351924

Applications and results of X-ray spectroscopy in implosion experiments on the National Ignition Facility
conference, January 2017

  • Epstein, R.; Regan, S. P.; Hammel, B. A.
  • ATOMIC PROCESSES IN PLASMAS APIP 2016: Proceedings of the 18th and 19th International Conference on Atomic Processes in Plasmas, AIP Conference Proceedings
  • DOI: 10.1063/1.4975747

Imaging of high-energy x-ray emission from cryogenic thermonuclear fuel implosions on the NIF
journal, October 2012

  • Ma, T.; Izumi, N.; Tommasini, R.
  • Review of Scientific Instruments, Vol. 83, Issue 10
  • DOI: 10.1063/1.4733313

X-ray and neutron sensitivity of imaging plates
conference, September 2013

  • Izumi, N.; Lee, J.; Romano, E.
  • SPIE Optical Engineering + Applications, SPIE Proceedings
  • DOI: 10.1117/12.2024513

Evaluation of the sensitivity and fading characteristics of an image plate system for x-ray diagnostics
journal, November 2008

  • Meadowcroft, A. L.; Bentley, C. D.; Stott, E. N.
  • Review of Scientific Instruments, Vol. 79, Issue 11
  • DOI: 10.1063/1.3013123

Design and performance of an imaging plate system for X-ray diffraction study
journal, April 1988

  • Amemiya, Yoshiyuki; Matsushita, Tadashi; Nakagawa, Atsushi
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 266, Issue 1-3
  • DOI: 10.1016/0168-9002(88)90458-5

Calibration and equivalency analysis of image plate scanners
journal, November 2014

  • Williams, G. Jackson; Maddox, Brian R.; Chen, Hui
  • Review of Scientific Instruments, Vol. 85, Issue 11
  • DOI: 10.1063/1.4886390

High-energy x-ray backlighter spectrum measurements using calibrated image plates
journal, February 2011

  • Maddox, B. R.; Park, H. S.; Remington, B. A.
  • Review of Scientific Instruments, Vol. 82, Issue 2
  • DOI: 10.1063/1.3531979

Calibrating image plate sensitivity in the 700 to 5000 eV spectral energy range
conference, September 2013

  • Haugh, Michael J.; Lee, Joshua; Romano, Edward
  • SPIE Optical Engineering + Applications, SPIE Proceedings
  • DOI: 10.1117/12.2024889

The quantum efficiency of radiographic imaging with image plates
journal, August 1996

  • Thoms, M.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 378, Issue 3
  • DOI: 10.1016/0168-9002(96)00530-x

Measurement of electron temperature of imploded capsules at the National Ignition Facility
journal, October 2012

  • Izumi, N.; Ma, T.; Barrios, M.
  • Review of Scientific Instruments, Vol. 83, Issue 10
  • DOI: 10.1063/1.4738660

    Works referencing / citing this record:

    X-ray spectrometer throughput model for (selected) flat Bragg crystal spectrometers on laser plasma facilities
    journal, October 2018

    • Thorn, D. B.; Coppari, F.; Döppner, T.
    • Review of Scientific Instruments, Vol. 89, Issue 10
    • DOI: 10.1063/1.5039423

    Image-plate sensitivity to x rays at 2 to 60 keV
    journal, January 2019

    • Rosenberg, M. J.; Thorn, D. B.; Izumi, N.
    • Review of Scientific Instruments, Vol. 90, Issue 1
    • DOI: 10.1063/1.5053592

    The Crystal Backlighter Imager: A spherically bent crystal imager for radiography on the National Ignition Facility
    journal, January 2019

    • Hall, G. N.; Krauland, C. M.; Schollmeier, M. S.
    • Review of Scientific Instruments, Vol. 90, Issue 1
    • DOI: 10.1063/1.5058700