Title: Readout models for BaFBr 0.85 I 0.15 :Eu image plates

The linearity of the photostimulated luminescence process makes repeated image-plate scanning a viable technique to extract more dynamic range. In order to obtain a response estimate, two semi-empirical models for the readout fading of an image plate are introduced; they relate the depth distribution of activated photostimulated luminescence centers within an image plate to the recorded signal. Model parameters are estimated from image-plate scan series with BAS-MS image plates and the Typhoon FLA 7000 scanner for the hard x-ray image-plate diagnostic over a collection of experiments providing x-ray energy spectra whose approximate shape is a double exponential.
Authors:
ORCiD logo [1] ;  [1]
  1. Univ. of Rochester, NY (United States). Lab. for Laser Energetics
Publication Date:
Grant/Contract Number:
NA0001944
Type:
Accepted Manuscript
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 89; Journal Issue: 6; Journal ID: ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Research Org:
Univ. of Rochester, NY (United States). Lab. for Laser Energetics
Sponsoring Org:
USDOE National Nuclear Security Administration (NNSA)
Country of Publication:
United States
Language:
English
Subject:
47 OTHER INSTRUMENTATION; Image plate; x-ray imaging; x-ray spectrometers; x-ray detectors; x-ray spectra