Abstract

The issue of detecting the XAFS signal from dilute samples is discussed in detail with the aim of making best use of high flux beamlines that provide up to 10(13)photonss(-1). Various detection methods are compared, including filters with slits, solid state detectors, crystal analyzers and combinations of these. These comparisons rely on simulations that use experimentally determined parameters. It is found that inelastic scattering places a fundamental limit on detection, and that it is important to take proper account of the polarization dependence of the signals. The combination of a filter-slit system with a solid state detector is a promising approach. With an optimized system good performance can be obtained even if the total count rate is limited to 10(7)Hz. Detection schemes with better energy resolution can help at the largest dilutions if their collection efficiency and count rate limits can be improved.

Authors:
 [1]
  1. Argonne National Lab. (ANL), Argonne, IL (United States). Advanced Photon Source (APS), X-ray Science Division
Publication Date:
Research Org.:
Argonne National Laboratory (ANL), Argonne, IL (United States)
Sponsoring Org.:
USDOE Office of Science (SC), Basic Energy Sciences (BES)
OSTI Identifier:
1261156
Alternate Identifier(s):
OSTI ID: 1396024
Grant/Contract Number:  
AC02-06CH11357
Resource Type:
Accepted Manuscript
Journal Name:
Journal of Synchrotron Radiation (Online)
Additional Journal Information:
Journal Name: Journal of Synchrotron Radiation (Online); Journal Volume: 22; Journal Issue: 2; Journal ID: ISSN 1600-5775
Publisher:
International Union of Crystallography
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; XAFS detectors; fluorescence; detection strategies; Detector; XAFS

Citation Formats

Heald, Steve M. Strategies and limitations for fluorescence detection of XAFS at high flux beamlines. United States: N. p., 2015. Web. doi:10.1107/s1600577515001320.
Heald, Steve M. Strategies and limitations for fluorescence detection of XAFS at high flux beamlines. United States. https://doi.org/10.1107/s1600577515001320
Heald, Steve M. Tue . "Strategies and limitations for fluorescence detection of XAFS at high flux beamlines". United States. https://doi.org/10.1107/s1600577515001320. https://www.osti.gov/servlets/purl/1261156.
@article{osti_1261156,
title = {Strategies and limitations for fluorescence detection of XAFS at high flux beamlines},
author = {Heald, Steve M.},
abstractNote = {The issue of detecting the XAFS signal from dilute samples is discussed in detail with the aim of making best use of high flux beamlines that provide up to 10(13)photonss(-1). Various detection methods are compared, including filters with slits, solid state detectors, crystal analyzers and combinations of these. These comparisons rely on simulations that use experimentally determined parameters. It is found that inelastic scattering places a fundamental limit on detection, and that it is important to take proper account of the polarization dependence of the signals. The combination of a filter-slit system with a solid state detector is a promising approach. With an optimized system good performance can be obtained even if the total count rate is limited to 10(7)Hz. Detection schemes with better energy resolution can help at the largest dilutions if their collection efficiency and count rate limits can be improved.},
doi = {10.1107/s1600577515001320},
journal = {Journal of Synchrotron Radiation (Online)},
number = 2,
volume = 22,
place = {United States},
year = {Tue Feb 17 00:00:00 EST 2015},
month = {Tue Feb 17 00:00:00 EST 2015}
}

Journal Article:
Free Publicly Available Full Text
Publisher's Version of Record

Citation Metrics:
Cited by: 19 works
Citation information provided by
Web of Science

Save / Share:

Works referenced in this record:

A broadband superconducting detector suitable for use in large arrays
journal, October 2003

  • Day, Peter K.; LeDuc, Henry G.; Mazin, Benjamin A.
  • Nature, Vol. 425, Issue 6960
  • DOI: 10.1038/nature02037

Filtered Energy Dispersive Detector (EDD) arrays: Superior detectors for EXAFS from very diluted solutions
journal, May 1986

  • K. Warburton, William
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 246, Issue 1-3
  • DOI: 10.1016/0168-9002(86)90148-8

Synchrotron X-ray fluorescence microprobe: A microanalytical instrument for trace element studies in geochemistry, cosmochemistry, and the soil and environmental sciences
journal, August 1994

  • Sutton, S. R.; Rivers, M. L.; Bajt, S.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 347, Issue 1-3
  • DOI: 10.1016/0168-9002(94)91919-4

Kilopixel X-ray Microcalorimeter Arrays for Astrophysics: Device Performance and Uniformity
journal, January 2012

  • Eckart, M. E.; Adams, J. S.; Bailey, C. N.
  • Journal of Low Temperature Physics, Vol. 167, Issue 5-6
  • DOI: 10.1007/s10909-012-0514-x

A plastic miniature x-ray emission spectrometer based on the cylindrical von Hamos geometry
journal, February 2012

  • Mattern, B. A.; Seidler, G. T.; Haave, M.
  • Review of Scientific Instruments, Vol. 83, Issue 2
  • DOI: 10.1063/1.3680598

Wavelength‐dispersive x‐ray fluorescence detector
journal, February 1995

  • Kirkland, J. P.; Kovantsev, V. E.; Dozier, C. M.
  • Review of Scientific Instruments, Vol. 66, Issue 2
  • DOI: 10.1063/1.1145924

Curved‐crystal (LiF) x‐ray focussing array for fluorescence EXAFS in dilute samples
journal, August 1980

  • Marcus, M.; Powers, L. S.; Storm, A. R.
  • Review of Scientific Instruments, Vol. 51, Issue 8
  • DOI: 10.1063/1.1136373

Wavelength-dispersive spectrometer for X-ray microfluorescence analysis at the X-ray microscopy beamline ID21 (ESRF)
journal, April 2010


Design and testing of a prototype multilayer analyzer X-ray fluorescence detector
journal, May 1999


X‐ray detection using a superconducting transition‐edge sensor microcalorimeter with electrothermal feedback
journal, September 1996

  • Irwin, K. D.; Hilton, G. C.; Wollman, D. A.
  • Applied Physics Letters, Vol. 69, Issue 13
  • DOI: 10.1063/1.117630

Silicon drift detector – the key to new experiments
journal, November 1998

  • Strüder, L.; Lechner, P.; Leutenegger, P.
  • The Science of Nature, Vol. 85, Issue 11
  • DOI: 10.1007/s001140050545

Local-Structure Determination at High Dilution: Internal Oxidation of 75-ppm Fe in Cu
journal, December 1979


X‐ray filter assembly for fluorescence measurements of x‐ray absorption fine structure
journal, December 1979

  • Stern, Edward A.; Heald, Steve M.
  • Review of Scientific Instruments, Vol. 50, Issue 12
  • DOI: 10.1063/1.1135763

Log spiral of revolution highly oriented pyrolytic graphite monochromator for fluorescence x-ray absorption edge fine structure
journal, September 2000

  • Pease, D. M.; Daniel, M.; Budnick, J. I.
  • Review of Scientific Instruments, Vol. 71, Issue 9
  • DOI: 10.1063/1.1287753

A bent Laue analyzer for fluorescence EXAFS detection
journal, May 1999


A 13-element Ge detector for fluorescence EXAFS
journal, April 1988

  • Cramer, S. P.; Tench, O.; Yocum, M.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 266, Issue 1-3
  • DOI: 10.1016/0168-9002(88)90449-4

Fluorescence detection of exafs: Sensitivity enhancement for dilute species and thin films
journal, September 1977


Recent tests of x-ray spectrometers using polycapillary optics
conference, October 2012

  • Heald, Steve; Seidler, Gerald T.; Mortensen, Devon
  • SPIE Optical Engineering + Applications, SPIE Proceedings
  • DOI: 10.1117/12.929960

Bent silicon crystal in the Laue geometry to resolve x-ray fluorescence for x-ray absorption spectroscopy
journal, November 2003

  • Kropf, A. J.; Finch, R. J.; Fortner, J. A.
  • Review of Scientific Instruments, Vol. 74, Issue 11
  • DOI: 10.1063/1.1618014

A short working distance multiple crystal x-ray spectrometer
journal, December 2008

  • Dickinson, B.; Seidler, G. T.; Webb, Z. W.
  • Review of Scientific Instruments, Vol. 79, Issue 12
  • DOI: 10.1063/1.3048544

An extraction algorithm for core-level excitations in non-resonant inelastic X-ray scattering spectra
journal, February 2008

  • Sternemann, H.; Sternemann, C.; Seidler, G. T.
  • Journal of Synchrotron Radiation, Vol. 15, Issue 2
  • DOI: 10.1107/S0909049508001696

Works referencing / citing this record:

ARDESIA: A fast silicon drift detector X-ray spectrometer for synchrotron applications
journal, February 2019

  • Hafizh, Idham; Bellotti, Giovanni; Carminati, Marco
  • X-Ray Spectrometry, Vol. 48, Issue 5
  • DOI: 10.1002/xrs.3017

2016 Atomic Spectrometry Update – a review of advances in X-ray fluorescence spectrometry and its applications
journal, January 2016

  • West, Margaret; Ellis, Andrew T.; Potts, Philip J.
  • Journal of Analytical Atomic Spectrometry, Vol. 31, Issue 9
  • DOI: 10.1039/c6ja90034h

Sub-ppm level high energy resolution fluorescence detected X-ray absorption spectroscopy of selenium in articular cartilage
journal, January 2019

  • Bissardon, C.; Proux, O.; Bureau, S.
  • The Analyst, Vol. 144, Issue 11
  • DOI: 10.1039/c9an00207c

Synchrotron hard X-ray chemical imaging of trace element speciation in heterogeneous samples: development of criteria for uncertainty analysis
journal, January 2020

  • Wielinski, Jonas; Marafatto, Francesco Femi; Gogos, Alexander
  • Journal of Analytical Atomic Spectrometry, Vol. 35, Issue 3
  • DOI: 10.1039/c9ja00394k

Improving the quality of XAFS data
journal, May 2018

  • Abe, Hitoshi; Aquilanti, Giuliana; Boada, Roberto
  • Journal of Synchrotron Radiation, Vol. 25, Issue 4
  • DOI: 10.1107/s1600577518006021

High-Energy Resolution Fluorescence Detected X-Ray Absorption Spectroscopy: A Powerful New Structural Tool in Environmental Biogeochemistry Sciences
journal, November 2017

  • Proux, Olivier; Lahera, Eric; Del Net, William
  • Journal of Environmental Quality, Vol. 46, Issue 6
  • DOI: 10.2134/jeq2017.01.0023