Title: Responsivity calibration of the LoWEUS spectrometer

Abstract

We performed an in situ calibration of the relative responsivity function of the Long-Wavelength Extreme Ultraviolet Spectrometer (LoWEUS), while operating on the Lithium Tokamak Experiment (LTX) at Princeton Plasma Physics Laboratory. The calibration was accomplished by measuring oxygen lines, which are typically present in LTX plasmas. The measured spectral line intensities of each oxygen charge state were then compared to the calculated emission strengths given in the CHIANTI atomic database. Normalizing the strongest line in each charge state to the CHIANTI predictions, we obtained the differences between the measured and predicted values for the relative strengths of the other lines of a given charge state. We find that a 3rd degree polynomial function provides a good fit to the data points. Lastly, our measurements show that the responsivity between about 120 and 300 Å varies by factor of ~30.

Authors:
 [1]; ORCiD logo [2];  [3];  [3]; ORCiD logo [3]
  1. Univ. of California, Berkeley, CA (United States)
  2. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
  3. Princeton Plasma Physics Lab. (PPPL), Princeton, NJ (United States)
Publication Date:
Research Org.:
Princeton Plasma Physics Lab. (PPPL), Princeton, NJ (United States)
Sponsoring Org.:
USDOE
OSTI Identifier:
1349666
Grant/Contract Number:  
AC02-09CH11466
Resource Type:
Accepted Manuscript
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 87; Journal Issue: 11; Journal ID: ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY

Citation Formats

Lepson, J. K., Beiersdorfer, P., Kaita, R., Majeski, R., and Boyle, D. Responsivity calibration of the LoWEUS spectrometer. United States: N. p., 2016. Web. doi:10.1063/1.4960729.
Lepson, J. K., Beiersdorfer, P., Kaita, R., Majeski, R., & Boyle, D. Responsivity calibration of the LoWEUS spectrometer. United States. doi:10.1063/1.4960729.
Lepson, J. K., Beiersdorfer, P., Kaita, R., Majeski, R., and Boyle, D. Fri . "Responsivity calibration of the LoWEUS spectrometer". United States. doi:10.1063/1.4960729. https://www.osti.gov/servlets/purl/1349666.
@article{osti_1349666,
title = {Responsivity calibration of the LoWEUS spectrometer},
author = {Lepson, J. K. and Beiersdorfer, P. and Kaita, R. and Majeski, R. and Boyle, D.},
abstractNote = {We performed an in situ calibration of the relative responsivity function of the Long-Wavelength Extreme Ultraviolet Spectrometer (LoWEUS), while operating on the Lithium Tokamak Experiment (LTX) at Princeton Plasma Physics Laboratory. The calibration was accomplished by measuring oxygen lines, which are typically present in LTX plasmas. The measured spectral line intensities of each oxygen charge state were then compared to the calculated emission strengths given in the CHIANTI atomic database. Normalizing the strongest line in each charge state to the CHIANTI predictions, we obtained the differences between the measured and predicted values for the relative strengths of the other lines of a given charge state. We find that a 3rd degree polynomial function provides a good fit to the data points. Lastly, our measurements show that the responsivity between about 120 and 300 Å varies by factor of ~30.},
doi = {10.1063/1.4960729},
journal = {Review of Scientific Instruments},
number = 11,
volume = 87,
place = {United States},
year = {2016},
month = {9}
}

Journal Article:
Free Publicly Available Full Text
Publisher's Version of Record

Citation Metrics:
Cited by: 2 works
Citation information provided by
Web of Science

Save / Share:

Works referenced in this record:

CHIANTI - an atomic database for emission lines: I. Wavelengths greater than 50 Å
journal, October 1997

  • Dere, K. P.; Landi, E.; Mason, H. E.
  • Astronomy and Astrophysics Supplement Series, Vol. 125, Issue 1
  • DOI: 10.1051/aas:1997368

High-resolution time-resolved extreme ultraviolet spectroscopy on NSTX
journal, October 2012

  • Lepson, J. K.; Beiersdorfer, P.; Clementson, J.
  • Review of Scientific Instruments, Vol. 83, Issue 10
  • DOI: 10.1063/1.4731753

Laboratory astrophysics in the extreme ultraviolet
journal, January 2008

  • Lepson, J. K.; Beiersdorfer, P.; Bitter, M.
  • Canadian Journal of Physics, Vol. 86, Issue 1
  • DOI: 10.1139/P07-167

Vacuum ultraviolet impurity spectroscopy on the Alcator C-Mod tokamak
journal, October 2010

  • Reinke, M. L.; Beiersdorfer, P.; Howard, N. T.
  • Review of Scientific Instruments, Vol. 81, Issue 10
  • DOI: 10.1063/1.3494380

Photometric calibration of an EUV flat field spectrometer at the advanced light source
journal, March 2003

  • May, M.; Lepson, J.; Beiersdorfer, P.
  • Review of Scientific Instruments, Vol. 74, Issue 3
  • DOI: 10.1063/1.1535247

Emission‐Line Spectra of S vii –S xiv in the 20–75 A Wavelength Region
journal, June 2005

  • Lepson, J. K.; Beiersdorfer, P.; Behar, E.
  • The Astrophysical Journal, Vol. 625, Issue 2
  • DOI: 10.1086/429879

Emission‐Line Spectra of Ar ix –Ar xvi in the Soft X‐Ray Region 20–50 A
journal, June 2003

  • Lepson, J. K.; Beiersdorfer, P.; Behar, E.
  • The Astrophysical Journal, Vol. 590, Issue 1
  • DOI: 10.1086/374980

EUV spectroscopy on NSTX
journal, July 2010

  • Lepson, J. K.; Beiersdorfer, P.; Clementson, J.
  • Journal of Physics B: Atomic, Molecular and Optical Physics, Vol. 43, Issue 14
  • DOI: 10.1088/0953-4075/43/14/144018

Chianti—An Atomic Database for Emission Lines. Xiii. soft X-Ray Improvements and Other Changes
journal, January 2013


    Works referencing / citing this record:

    Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI
    journal, February 2018

    • Weller, M. E.; Beiersdorfer, P.; Soukhanovskii, V. A.
    • The Astrophysical Journal, Vol. 854, Issue 2
    • DOI: 10.3847/1538-4357/aaa5a1

    Electron-density-sensitive Line Ratios of Fe xiii– xvi from Laboratory Sources Compared to CHIANTI
    journal, February 2018

    • Weller, M. E.; Beiersdorfer, P.; Soukhanovskii, V. A.
    • The Astrophysical Journal, Vol. 854, Issue 2
    • DOI: 10.3847/1538-4357/aaa5a1