Procedure for field axes measurement, beam indication adjustment, and figure of convergence determination within performance tests for radiation therapy equipment

Abstract

A routine measuring procedure for the verification of radiation field axes and figure of convergence within a spatial resolution of +- 0.5 mm is described. Measurements are done in two parallel planes in a certain distance before and behind the presumed isocentre. The used test arrangement permits rapid check and controlled adjustment of the alignment of beam or isocentre indicating devices for all isocentric radiation therapy equipment.
Authors:
Quast, U; Krause, K; Rassow, J [1] 
  1. Essen Univ. (Gesamthochschule) (Germany, F.R.). Abt. fuer Klinische Strahlenphysik
Publication Date:
Sep 01, 1976
Product Type:
Journal Article
Reference Number:
AIX-08-281465; EDB-77-038010
Resource Relation:
Journal Name: Strahlentherapie; (Germany, Federal Republic of); Journal Volume: 152:3; Other Information: 2 figs
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 07 ISOTOPES AND RADIATION SOURCES; BEAM MONITORS; RADIOTHERAPY; EQUIPMENT; ALIGNMENT; BEAM OPTICS; DISTANCE; PERFORMANCE TESTING; SIMULATION; SPATIAL DOSE DISTRIBUTIONS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; TESTING; THERAPY; 440101* - Radiation Instrumentation- General Detectors or Monitors & Radiometric Instruments; 070200 - Radiation Sources
OSTI ID:
7336370
Country of Origin:
Germany
Language:
English
Other Identifying Numbers:
Journal ID: CODEN: STRAA
Submitting Site:
INIS
Size:
Pages: 268-272
Announcement Date:
Mar 01, 1977

Citation Formats

Quast, U, Krause, K, and Rassow, J. Procedure for field axes measurement, beam indication adjustment, and figure of convergence determination within performance tests for radiation therapy equipment. Germany: N. p., 1976. Web.
Quast, U, Krause, K, & Rassow, J. Procedure for field axes measurement, beam indication adjustment, and figure of convergence determination within performance tests for radiation therapy equipment. Germany.
Quast, U, Krause, K, and Rassow, J. 1976. "Procedure for field axes measurement, beam indication adjustment, and figure of convergence determination within performance tests for radiation therapy equipment." Germany.
@misc{etde_7336370,
title = {Procedure for field axes measurement, beam indication adjustment, and figure of convergence determination within performance tests for radiation therapy equipment}
author = {Quast, U, Krause, K, and Rassow, J}
abstractNote = {A routine measuring procedure for the verification of radiation field axes and figure of convergence within a spatial resolution of +- 0.5 mm is described. Measurements are done in two parallel planes in a certain distance before and behind the presumed isocentre. The used test arrangement permits rapid check and controlled adjustment of the alignment of beam or isocentre indicating devices for all isocentric radiation therapy equipment.}
journal = []
volume = {152:3}
journal type = {AC}
place = {Germany}
year = {1976}
month = {Sep}
}