Abstract
By this study, a graphite calorimeter has been developed to measure the average ion energy from electrostatic ion accelerator (3 MV Tandem Ion Accelerator of the Takasaki Ion Accelerators for Advanced Radiation Application). A graphite calorimeter was designed for measuring the ion beam energy in wide ion energy ranges (from 3 MeV to 10 MeV). By a couple of this study, the {sup 1} H({sup 15} N, {alpha} {gamma}) {sup 12} C resonant nuclear reactions (resonant energy 6.385 MeV and 13.355 MeV) has been applied for measuring the absolute ion beam energy serve as a comparative method. The obtained results were compared in terms of resolution and accuracy. Close agreement was found for both methods, the deviation was observed within {+-} 3%
Citation Formats
Bunnak, S, Mizuhashi, K, and Tajima, S.
Energy calibration of the 3 MV tandem accelerator.
Thailand: N. p.,
1996.
Web.
Bunnak, S, Mizuhashi, K, & Tajima, S.
Energy calibration of the 3 MV tandem accelerator.
Thailand.
Bunnak, S, Mizuhashi, K, and Tajima, S.
1996.
"Energy calibration of the 3 MV tandem accelerator."
Thailand.
@misc{etde_635368,
title = {Energy calibration of the 3 MV tandem accelerator}
author = {Bunnak, S, Mizuhashi, K, and Tajima, S}
abstractNote = {By this study, a graphite calorimeter has been developed to measure the average ion energy from electrostatic ion accelerator (3 MV Tandem Ion Accelerator of the Takasaki Ion Accelerators for Advanced Radiation Application). A graphite calorimeter was designed for measuring the ion beam energy in wide ion energy ranges (from 3 MeV to 10 MeV). By a couple of this study, the {sup 1} H({sup 15} N, {alpha} {gamma}) {sup 12} C resonant nuclear reactions (resonant energy 6.385 MeV and 13.355 MeV) has been applied for measuring the absolute ion beam energy serve as a comparative method. The obtained results were compared in terms of resolution and accuracy. Close agreement was found for both methods, the deviation was observed within {+-} 3%}
place = {Thailand}
year = {1996}
month = {Dec}
}
title = {Energy calibration of the 3 MV tandem accelerator}
author = {Bunnak, S, Mizuhashi, K, and Tajima, S}
abstractNote = {By this study, a graphite calorimeter has been developed to measure the average ion energy from electrostatic ion accelerator (3 MV Tandem Ion Accelerator of the Takasaki Ion Accelerators for Advanced Radiation Application). A graphite calorimeter was designed for measuring the ion beam energy in wide ion energy ranges (from 3 MeV to 10 MeV). By a couple of this study, the {sup 1} H({sup 15} N, {alpha} {gamma}) {sup 12} C resonant nuclear reactions (resonant energy 6.385 MeV and 13.355 MeV) has been applied for measuring the absolute ion beam energy serve as a comparative method. The obtained results were compared in terms of resolution and accuracy. Close agreement was found for both methods, the deviation was observed within {+-} 3%}
place = {Thailand}
year = {1996}
month = {Dec}
}