You need JavaScript to view this

Energy calibration of the 3 MV tandem accelerator

Abstract

By this study, a graphite calorimeter has been developed to measure the average ion energy from electrostatic ion accelerator (3 MV Tandem Ion Accelerator of the Takasaki Ion Accelerators for Advanced Radiation Application). A graphite calorimeter was designed for measuring the ion beam energy in wide ion energy ranges (from 3 MeV to 10 MeV). By a couple of this study, the {sup 1} H({sup 15} N, {alpha} {gamma}) {sup 12} C resonant nuclear reactions (resonant energy 6.385 MeV and 13.355 MeV) has been applied for measuring the absolute ion beam energy serve as a comparative method. The obtained results were compared in terms of resolution and accuracy. Close agreement was found for both methods, the deviation was observed within {+-} 3%
Publication Date:
Dec 01, 1996
Product Type:
Miscellaneous
Report Number:
INIS-TH-006; CONF-9612121-
Reference Number:
SCA: 430303; 430100; PA: AIX-29:044579; EDB-98:089199; SN: 98001995033
Resource Relation:
Conference: 6. conference on nuclear science and technology, Bangkok (Thailand), 2-4 Dec 1996; Other Information: PBD: Dec 1996; Related Information: Is Part Of Proceedings of the 6th nuclear science and technology conference; PB: 761 p.
Subject:
43 PARTICLE ACCELERATORS; CALIBRATION; CALORIMETERS; EXPERIMENTAL DATA; TANDEM ELECTROSTATIC ACCELERATORS
OSTI ID:
635368
Research Organizations:
Office of Atomic Energy for Peace, Bangkok (Thailand)
Country of Origin:
Thailand
Language:
Thai
Other Identifying Numbers:
Other: ON: DE98632369; TRN: TH9800072044579
Availability:
INIS; OSTI as DE98632369
Submitting Site:
INIS
Size:
pp. [16]
Announcement Date:
Sep 09, 1998

Citation Formats

Bunnak, S, Mizuhashi, K, and Tajima, S. Energy calibration of the 3 MV tandem accelerator. Thailand: N. p., 1996. Web.
Bunnak, S, Mizuhashi, K, & Tajima, S. Energy calibration of the 3 MV tandem accelerator. Thailand.
Bunnak, S, Mizuhashi, K, and Tajima, S. 1996. "Energy calibration of the 3 MV tandem accelerator." Thailand.
@misc{etde_635368,
title = {Energy calibration of the 3 MV tandem accelerator}
author = {Bunnak, S, Mizuhashi, K, and Tajima, S}
abstractNote = {By this study, a graphite calorimeter has been developed to measure the average ion energy from electrostatic ion accelerator (3 MV Tandem Ion Accelerator of the Takasaki Ion Accelerators for Advanced Radiation Application). A graphite calorimeter was designed for measuring the ion beam energy in wide ion energy ranges (from 3 MeV to 10 MeV). By a couple of this study, the {sup 1} H({sup 15} N, {alpha} {gamma}) {sup 12} C resonant nuclear reactions (resonant energy 6.385 MeV and 13.355 MeV) has been applied for measuring the absolute ion beam energy serve as a comparative method. The obtained results were compared in terms of resolution and accuracy. Close agreement was found for both methods, the deviation was observed within {+-} 3%}
place = {Thailand}
year = {1996}
month = {Dec}
}