Abstract
In this paper the commercially available bubble neutron detectors (bubble damage neutron detectors (BDNDs*) from Bubble Technology Industries, Chalk River; and superheated drop detectors (SDDs*) from APFEL Industries, New Haven) for lowest limit of detection of an individual neutron dosimeter were tested. They were tested with the different neutron sources. BDNDs* tested had the sensitivity about 1 bubble per 1 Sv of H*(10) of AmBe neutrons, they were evaluated by eye counting (20 to 30 bubbles per detector). Two types of reusable BDNDs* were tested: BD-100R without and with temperature compensation, both with neutron energy threshold about 100 keV. SDDs* tested had the sensitivity about 3 bubbles per 1 {mu}Sv of H*10 from AmBe neutrons, they were evaluated using APFEL Survey Meter Model 202. SDDs* with three different energy thresholds have been used: 0.1, 1 and 6 MeV. For energetical dependence of BDNDs* the general conclusions were formulated in the following way: (1) With the exception of thermal neutron source SIGMA (50% of H*(10) from thermal neutrons) and high energy reference fields there is a reasonable agreement of data measured with BDNDs* and expected values; (2) the new lots to have a little different energetic dependence. The relative responses for
More>>
Spurny, F;
Vlcek, B
[1]
- Academy of the Sciences of Czech Republic, Prague (Czech Republic). Nuclear Physics Institute, Department of Radiation Dosimetry
Citation Formats
Spurny, F, and Vlcek, B.
To the use of bubble detectors in personal neutron dosimetry.
Slovakia: N. p.,
1995.
Web.
Spurny, F, & Vlcek, B.
To the use of bubble detectors in personal neutron dosimetry.
Slovakia.
Spurny, F, and Vlcek, B.
1995.
"To the use of bubble detectors in personal neutron dosimetry."
Slovakia.
@misc{etde_551056,
title = {To the use of bubble detectors in personal neutron dosimetry}
author = {Spurny, F, and Vlcek, B}
abstractNote = {In this paper the commercially available bubble neutron detectors (bubble damage neutron detectors (BDNDs*) from Bubble Technology Industries, Chalk River; and superheated drop detectors (SDDs*) from APFEL Industries, New Haven) for lowest limit of detection of an individual neutron dosimeter were tested. They were tested with the different neutron sources. BDNDs* tested had the sensitivity about 1 bubble per 1 Sv of H*(10) of AmBe neutrons, they were evaluated by eye counting (20 to 30 bubbles per detector). Two types of reusable BDNDs* were tested: BD-100R without and with temperature compensation, both with neutron energy threshold about 100 keV. SDDs* tested had the sensitivity about 3 bubbles per 1 {mu}Sv of H*10 from AmBe neutrons, they were evaluated using APFEL Survey Meter Model 202. SDDs* with three different energy thresholds have been used: 0.1, 1 and 6 MeV. For energetical dependence of BDNDs* the general conclusions were formulated in the following way: (1) With the exception of thermal neutron source SIGMA (50% of H*(10) from thermal neutrons) and high energy reference fields there is a reasonable agreement of data measured with BDNDs* and expected values; (2) the new lots to have a little different energetic dependence. The relative responses for `soft` fields are for them systematically higher than for previous samples. The response to energies between 0.01 and 1 MeV is for these lots relatively higher. (3) The underestimation of high energy neutrons is typical for any LET-threshold type detectors.It should be kept in mind when BDNDs* are used as dosemeters in high energy neutron environment. For energetical dependence of SDDs* was concluded: (1) The energetical dependence of SDD 100 is comparable with the dependencies of BD-100R and PND, the underestimation of high energy neutrons included; (2) The use of SDD with different energy thresholds can provide interesting spectrometric information; (Abstract Truncated)}
place = {Slovakia}
year = {1995}
month = {Dec}
}
title = {To the use of bubble detectors in personal neutron dosimetry}
author = {Spurny, F, and Vlcek, B}
abstractNote = {In this paper the commercially available bubble neutron detectors (bubble damage neutron detectors (BDNDs*) from Bubble Technology Industries, Chalk River; and superheated drop detectors (SDDs*) from APFEL Industries, New Haven) for lowest limit of detection of an individual neutron dosimeter were tested. They were tested with the different neutron sources. BDNDs* tested had the sensitivity about 1 bubble per 1 Sv of H*(10) of AmBe neutrons, they were evaluated by eye counting (20 to 30 bubbles per detector). Two types of reusable BDNDs* were tested: BD-100R without and with temperature compensation, both with neutron energy threshold about 100 keV. SDDs* tested had the sensitivity about 3 bubbles per 1 {mu}Sv of H*10 from AmBe neutrons, they were evaluated using APFEL Survey Meter Model 202. SDDs* with three different energy thresholds have been used: 0.1, 1 and 6 MeV. For energetical dependence of BDNDs* the general conclusions were formulated in the following way: (1) With the exception of thermal neutron source SIGMA (50% of H*(10) from thermal neutrons) and high energy reference fields there is a reasonable agreement of data measured with BDNDs* and expected values; (2) the new lots to have a little different energetic dependence. The relative responses for `soft` fields are for them systematically higher than for previous samples. The response to energies between 0.01 and 1 MeV is for these lots relatively higher. (3) The underestimation of high energy neutrons is typical for any LET-threshold type detectors.It should be kept in mind when BDNDs* are used as dosemeters in high energy neutron environment. For energetical dependence of SDDs* was concluded: (1) The energetical dependence of SDD 100 is comparable with the dependencies of BD-100R and PND, the underestimation of high energy neutrons included; (2) The use of SDD with different energy thresholds can provide interesting spectrometric information; (Abstract Truncated)}
place = {Slovakia}
year = {1995}
month = {Dec}
}