Abstract
The calibration of an optical pyrometer against a tungsten strip lamp was extrapolated from the maximum temperature of the lamp, 2300/sup 0/C to 3100/sup 0/C. The accuracy of the extrapolation was shown to be within 0.65% by the use of sectored discs and a unique tungsten black-body source. 14 references, 2 figures.
Citation Formats
Hoff, H A, Rothman, S J, and Mundy, J N.
Calibration of an optical pyrometer above 2300/sup 0/ C.
United Kingdom: N. p.,
1985.
Web.
Hoff, H A, Rothman, S J, & Mundy, J N.
Calibration of an optical pyrometer above 2300/sup 0/ C.
United Kingdom.
Hoff, H A, Rothman, S J, and Mundy, J N.
1985.
"Calibration of an optical pyrometer above 2300/sup 0/ C."
United Kingdom.
@misc{etde_5183553,
title = {Calibration of an optical pyrometer above 2300/sup 0/ C}
author = {Hoff, H A, Rothman, S J, and Mundy, J N}
abstractNote = {The calibration of an optical pyrometer against a tungsten strip lamp was extrapolated from the maximum temperature of the lamp, 2300/sup 0/C to 3100/sup 0/C. The accuracy of the extrapolation was shown to be within 0.65% by the use of sectored discs and a unique tungsten black-body source. 14 references, 2 figures.}
journal = []
volume = {18}
journal type = {AC}
place = {United Kingdom}
year = {1985}
month = {Jan}
}
title = {Calibration of an optical pyrometer above 2300/sup 0/ C}
author = {Hoff, H A, Rothman, S J, and Mundy, J N}
abstractNote = {The calibration of an optical pyrometer against a tungsten strip lamp was extrapolated from the maximum temperature of the lamp, 2300/sup 0/C to 3100/sup 0/C. The accuracy of the extrapolation was shown to be within 0.65% by the use of sectored discs and a unique tungsten black-body source. 14 references, 2 figures.}
journal = []
volume = {18}
journal type = {AC}
place = {United Kingdom}
year = {1985}
month = {Jan}
}