Tests of a new CCD-camera based neutron radiography detector system at the reactor stations in Munich and Vienna

Abstract

The performance of the new neutron radiography detector designed at PSI with a cooled high sensitive CCD-camera was investigated under real neutronic conditions at three beam ports of two reactor stations. Different converter screens were applied for which the sensitivity and the modulation transfer function (MTF) could be obtained. The results are very encouraging concerning the utilization of this detector system as standard tool at the radiography stations at the spallation source SINQ. (author) 3 figs., 5 refs.
Authors:
Lehmann, E; Pleinert, H; [1]  Schillinger, B; [2]  Koerner, S [3] 
  1. Paul Scherrer Inst. (PSI), Villigen (Switzerland)
  2. Technische Univ. Muenchen (Germany)
  3. Atominstitut der Oesterreichischen Universitaeten, Vienna (Austria)
Publication Date:
Sep 01, 1997
Product Type:
Miscellaneous
Report Number:
INIS-CH-004
Reference Number:
SCA: 440101; PA: AIX-28:057441; EDB-97:120068; SN: 97001841663
Resource Relation:
Other Information: PBD: [1997]; Related Information: Is Part Of Paul Scherrer Institut annual report 1996. Annex IIIA: solid state research at large facilities; Baltensperger, U.; Lorenzen, R. [eds.] [Paul Scherrer Inst. (PSI), Villigen (Switzerland)]; PB: 180 p.
Subject:
44 INSTRUMENTATION, INCLUDING NUCLEAR AND PARTICLE DETECTORS; NEUTRON CAMERAS; NEUTRON RADIOGRAPHY; NEUTRON DETECTORS; NEUTRON SOURCES; SPALLATION; SWITZERLAND; EXPERIMENTAL DATA; FRM REACTOR; THEORETICAL DATA; TRIGA-2-VIENNA REACTOR
OSTI ID:
517360
Country of Origin:
Switzerland
Language:
English
Other Identifying Numbers:
Other: ON: DE97638271; TRN: CH9700444057441
Availability:
OSTI as DE97638271
Submitting Site:
CHN
Size:
pp. 15-16
Announcement Date:
Sep 22, 1997

Citation Formats

Lehmann, E, Pleinert, H, Schillinger, B, and Koerner, S. Tests of a new CCD-camera based neutron radiography detector system at the reactor stations in Munich and Vienna. Switzerland: N. p., 1997. Web.
Lehmann, E, Pleinert, H, Schillinger, B, & Koerner, S. Tests of a new CCD-camera based neutron radiography detector system at the reactor stations in Munich and Vienna. Switzerland.
Lehmann, E, Pleinert, H, Schillinger, B, and Koerner, S. 1997. "Tests of a new CCD-camera based neutron radiography detector system at the reactor stations in Munich and Vienna." Switzerland.
@misc{etde_517360,
title = {Tests of a new CCD-camera based neutron radiography detector system at the reactor stations in Munich and Vienna}
author = {Lehmann, E, Pleinert, H, Schillinger, B, and Koerner, S}
abstractNote = {The performance of the new neutron radiography detector designed at PSI with a cooled high sensitive CCD-camera was investigated under real neutronic conditions at three beam ports of two reactor stations. Different converter screens were applied for which the sensitivity and the modulation transfer function (MTF) could be obtained. The results are very encouraging concerning the utilization of this detector system as standard tool at the radiography stations at the spallation source SINQ. (author) 3 figs., 5 refs.}
place = {Switzerland}
year = {1997}
month = {Sep}
}