Abstract
The chloride present in the MOX fuel is separated from the matrix by pyrohydrolysis at a temperature of 950 {+-} 50 degC and is then analyzed by chloride ion selective electrode (Cl-ISE). The range covered is 0.4-4 ppm with a precision of better than {+-}5% R.S.D. (author). 4 refs., 1 tab.
Govindan, R;
Das, D K;
Mallik, G K;
Sumathi, A;
Patil, Sangeeta;
Raul, Seema;
Bhargava, V K;
Kamath, H S
[1]
- Bhabha Atomic Research Centre, Tarapur (India). Advanced Fuel Fabrication Facility
Citation Formats
Govindan, R, Das, D K, Mallik, G K, Sumathi, A, Patil, Sangeeta, Raul, Seema, Bhargava, V K, and Kamath, H S.
Determination of chloride in MOX samples using chloride ion selective electrode.
India: N. p.,
1997.
Web.
Govindan, R, Das, D K, Mallik, G K, Sumathi, A, Patil, Sangeeta, Raul, Seema, Bhargava, V K, & Kamath, H S.
Determination of chloride in MOX samples using chloride ion selective electrode.
India.
Govindan, R, Das, D K, Mallik, G K, Sumathi, A, Patil, Sangeeta, Raul, Seema, Bhargava, V K, and Kamath, H S.
1997.
"Determination of chloride in MOX samples using chloride ion selective electrode."
India.
@misc{etde_510705,
title = {Determination of chloride in MOX samples using chloride ion selective electrode}
author = {Govindan, R, Das, D K, Mallik, G K, Sumathi, A, Patil, Sangeeta, Raul, Seema, Bhargava, V K, and Kamath, H S}
abstractNote = {The chloride present in the MOX fuel is separated from the matrix by pyrohydrolysis at a temperature of 950 {+-} 50 degC and is then analyzed by chloride ion selective electrode (Cl-ISE). The range covered is 0.4-4 ppm with a precision of better than {+-}5% R.S.D. (author). 4 refs., 1 tab.}
place = {India}
year = {1997}
month = {Sep}
}
title = {Determination of chloride in MOX samples using chloride ion selective electrode}
author = {Govindan, R, Das, D K, Mallik, G K, Sumathi, A, Patil, Sangeeta, Raul, Seema, Bhargava, V K, and Kamath, H S}
abstractNote = {The chloride present in the MOX fuel is separated from the matrix by pyrohydrolysis at a temperature of 950 {+-} 50 degC and is then analyzed by chloride ion selective electrode (Cl-ISE). The range covered is 0.4-4 ppm with a precision of better than {+-}5% R.S.D. (author). 4 refs., 1 tab.}
place = {India}
year = {1997}
month = {Sep}
}