Radiation dose to the patient in several diagnostic X-ray examinations performed with conventional radiographic equipment in 3 major centers in Israel, a comparative study

Abstract

In this survey we determined the effective dose to the patient in specific x-ray diagnostic procedures performed in the radiology Department and emergency rooms in 3 major hospitals (A, B, and C) in Israel. In each hospital we measured the dose area product (DAP) in a number of several major diagnostic procedures. In each hospital we carried out measurements in two X-ray departments (A1, A2. B1, B2, C1, C2). The DAP measurements were carried out using a Diamentor Dose Area Product meter (DAP). The entrance dose was obtained using calibration measurements relating the exposure at a reference distance to the imaging parameters (kVp, mAs, filtration and distance) that were recorded in the examinations. The results will be presented and compared to the reference international guidance values (authors).
Authors:
Ben-Shlomo, A; Schlesinger, T; [1]  Kushilevsky, A [2] 
  1. Israel Atomic Energy Commission, Yavne (Israel). Soreq Nuclear Research Center
  2. Ben-Gurion Univ. of the Negev, Beersheba (Israel) Dept. of Biomedical Engineering
Publication Date:
Dec 01, 1996
Product Type:
Conference
Report Number:
INIS-mf-15507; CONF-961252-
Reference Number:
SCA: 560101; PA: AIX-28:023362; EDB-97:075946; SN: 97001788657
Resource Relation:
Conference: 19. conference of the Israel Nuclear Societies, Herzliya (Israel), 9-10 Dec 1996; Other Information: PBD: Dec 1996; Related Information: Is Part Of Program and book of abstracts; PB: 149 p.
Subject:
56 BIOLOGY AND MEDICINE, APPLIED STUDIES; X-RAY DOSIMETRY; PATIENTS; DIAGNOSTIC USES; RADIATION DOSES
OSTI ID:
475923
Research Organizations:
Israel Nuclear Society, Yavne (Israel)
Country of Origin:
Israel
Language:
English
Other Identifying Numbers:
Other: ON: DE97616259; TRN: IL9606180023362
Availability:
INIS; OSTI as DE97616259
Submitting Site:
INIS
Size:
pp. 1-4
Announcement Date:

Citation Formats

Ben-Shlomo, A, Schlesinger, T, and Kushilevsky, A. Radiation dose to the patient in several diagnostic X-ray examinations performed with conventional radiographic equipment in 3 major centers in Israel, a comparative study. Israel: N. p., 1996. Web.
Ben-Shlomo, A, Schlesinger, T, & Kushilevsky, A. Radiation dose to the patient in several diagnostic X-ray examinations performed with conventional radiographic equipment in 3 major centers in Israel, a comparative study. Israel.
Ben-Shlomo, A, Schlesinger, T, and Kushilevsky, A. 1996. "Radiation dose to the patient in several diagnostic X-ray examinations performed with conventional radiographic equipment in 3 major centers in Israel, a comparative study." Israel.
@misc{etde_475923,
title = {Radiation dose to the patient in several diagnostic X-ray examinations performed with conventional radiographic equipment in 3 major centers in Israel, a comparative study}
author = {Ben-Shlomo, A, Schlesinger, T, and Kushilevsky, A}
abstractNote = {In this survey we determined the effective dose to the patient in specific x-ray diagnostic procedures performed in the radiology Department and emergency rooms in 3 major hospitals (A, B, and C) in Israel. In each hospital we measured the dose area product (DAP) in a number of several major diagnostic procedures. In each hospital we carried out measurements in two X-ray departments (A1, A2. B1, B2, C1, C2). The DAP measurements were carried out using a Diamentor Dose Area Product meter (DAP). The entrance dose was obtained using calibration measurements relating the exposure at a reference distance to the imaging parameters (kVp, mAs, filtration and distance) that were recorded in the examinations. The results will be presented and compared to the reference international guidance values (authors).}
place = {Israel}
year = {1996}
month = {Dec}
}