Abstract
The present work describe the quality control procedures carried out on three RIA-Kits by the Isotopes Centre of Cuba, The subject matter of study were ;: were: LH-RIA-Kit, FSH-RIA-Kit and Prolactin- RIA -Kit. The controls have included the characterization of the {sup 125}I labelled hormones the specific antibodies, the 2 nd antibodies, the standards curves and the control serum. For the validation of these Kits were used reference standards from the WHO (World Health Organizations) and Kits from CIS company (France) based on the IRMA assays technologies . The results obtained allow us encourage the reliability of RIA-Kits
Caso Pena, R;
[1]
Arranz Calzado, C
[2]
- Centro de Isotopos, La Habana (Cuba)
- Instituto Nacional de Endocrinologia, La Habana (Cuba)
Citation Formats
Caso Pena, R, and Arranz Calzado, C.
Quality control of radioimmunoassay kits of pituitary hormones; Control de calidad de kits de radioinmunoanalisis de hormonas hipofisiarias.
Cuba: N. p.,
1997.
Web.
Caso Pena, R, & Arranz Calzado, C.
Quality control of radioimmunoassay kits of pituitary hormones; Control de calidad de kits de radioinmunoanalisis de hormonas hipofisiarias.
Cuba.
Caso Pena, R, and Arranz Calzado, C.
1997.
"Quality control of radioimmunoassay kits of pituitary hormones; Control de calidad de kits de radioinmunoanalisis de hormonas hipofisiarias."
Cuba.
@misc{etde_343486,
title = {Quality control of radioimmunoassay kits of pituitary hormones; Control de calidad de kits de radioinmunoanalisis de hormonas hipofisiarias}
author = {Caso Pena, R, and Arranz Calzado, C}
abstractNote = {The present work describe the quality control procedures carried out on three RIA-Kits by the Isotopes Centre of Cuba, The subject matter of study were ;: were: LH-RIA-Kit, FSH-RIA-Kit and Prolactin- RIA -Kit. The controls have included the characterization of the {sup 125}I labelled hormones the specific antibodies, the 2 nd antibodies, the standards curves and the control serum. For the validation of these Kits were used reference standards from the WHO (World Health Organizations) and Kits from CIS company (France) based on the IRMA assays technologies . The results obtained allow us encourage the reliability of RIA-Kits}
place = {Cuba}
year = {1997}
month = {Dec}
}
title = {Quality control of radioimmunoassay kits of pituitary hormones; Control de calidad de kits de radioinmunoanalisis de hormonas hipofisiarias}
author = {Caso Pena, R, and Arranz Calzado, C}
abstractNote = {The present work describe the quality control procedures carried out on three RIA-Kits by the Isotopes Centre of Cuba, The subject matter of study were ;: were: LH-RIA-Kit, FSH-RIA-Kit and Prolactin- RIA -Kit. The controls have included the characterization of the {sup 125}I labelled hormones the specific antibodies, the 2 nd antibodies, the standards curves and the control serum. For the validation of these Kits were used reference standards from the WHO (World Health Organizations) and Kits from CIS company (France) based on the IRMA assays technologies . The results obtained allow us encourage the reliability of RIA-Kits}
place = {Cuba}
year = {1997}
month = {Dec}
}