You need JavaScript to view this

Performance and beam characteristics of the PANTAK THERAPAX HF225 X-ray therapy machine

Abstract

The performance and beam characteristics of the new PANTAK THERAPAX HF225 X-ray therapy machine have been measured, evaluated and discussed. Eight beam qualities within the working range of generating potentials between 50 and 225 kVp are used in our department. These beam qualities have been investigated in order to provide a data base specific to our machine. Beam Quality, Central Axis Depth Dose, Output, Relative Field Uniformity and Timer Error were investigated. (authors) 11 refs., 4 figs., 9 tabs.
Authors:
Yiannakkaras, C; Papadopoulos, N; Christodoulides, G [1] 
  1. Department of Medical Physics, Nicosia General Hospital, 1450 Nicosia (Cyprus)
Publication Date:
Dec 31, 1998
Product Type:
Conference
Report Number:
INIS-CY-0004; CONF-9806202-
Reference Number:
SCA: 550603; PA: AIX-30:017099; EDB-99:043422; SN: 99002082142
Resource Relation:
Conference: 8. Mediterranean conference on medical and biological engineering and computing (Medicon `98), Lemesos (Cyprus), 14-17 Jun 1998; Other Information: DN: 11 refs., 4 figs., 9 tabs. The proceedings of this meeting are available on CD-ROM in PDF, at the price of US$50, from the Cyprus Association of Medical Physics and Biomedical Engineering, P.O.Box 24039, 1700 Nicosia, Cyprus. The CD-ROM makes use of Abode Acrobat PDF (Portable Document Format). To read the files the Abode Acrobat reader has to be installed on your system. A copy is provided with the CD-ROM; PBD: 1998; Related Information: Is Part Of Proceedings of the 8. Mediterranean Conference on Medical and Biological Engineering and Computing (Medicon `98); Christofides, Stelios; Pattichis, Constantinos; Schizas, Christos; Keravnou-Papailiou, Elpida; Kaplanis, Prodromos; Spyros, Spyrou; Christodoulides, George; Theodoulou, Yiannis [eds.]; PB: 220 p.
Subject:
55 BIOLOGY AND MEDICINE, BASIC STUDIES; BEAM PROFILES; DEPTH DOSE DISTRIBUTIONS; HALF-THICKNESS; ISODOSE CURVES; RADIATION DOSES; RADIATION QUALITY; RADIOTHERAPY; X RADIATION
OSTI ID:
331906
Research Organizations:
Cyprus Association of Medical Physics and Biomedical Engineering (CAMPBE), Nicosia (Cyprus); The Department of Computer Science, University of Cyprus, Nicosia (Cyprus)
Country of Origin:
Cyprus
Language:
English
Other Identifying Numbers:
Other: ON: DE99619912; ISBN 9963-607-14-4; TRN: CY9900022017099
Availability:
INIS; OSTI as DE99619912
Submitting Site:
INIS
Size:
pp. 133
Announcement Date:

Citation Formats

Yiannakkaras, C, Papadopoulos, N, and Christodoulides, G. Performance and beam characteristics of the PANTAK THERAPAX HF225 X-ray therapy machine. Cyprus: N. p., 1998. Web.
Yiannakkaras, C, Papadopoulos, N, & Christodoulides, G. Performance and beam characteristics of the PANTAK THERAPAX HF225 X-ray therapy machine. Cyprus.
Yiannakkaras, C, Papadopoulos, N, and Christodoulides, G. 1998. "Performance and beam characteristics of the PANTAK THERAPAX HF225 X-ray therapy machine." Cyprus.
@misc{etde_331906,
title = {Performance and beam characteristics of the PANTAK THERAPAX HF225 X-ray therapy machine}
author = {Yiannakkaras, C, Papadopoulos, N, and Christodoulides, G}
abstractNote = {The performance and beam characteristics of the new PANTAK THERAPAX HF225 X-ray therapy machine have been measured, evaluated and discussed. Eight beam qualities within the working range of generating potentials between 50 and 225 kVp are used in our department. These beam qualities have been investigated in order to provide a data base specific to our machine. Beam Quality, Central Axis Depth Dose, Output, Relative Field Uniformity and Timer Error were investigated. (authors) 11 refs., 4 figs., 9 tabs.}