Endovascular Irradiations with beta sources

Abstract

For treatment of restenoses tubes (inner/outer diameter 1 and 2 mm; length 3 or 5 mm) with Y-90 foils, shielded by Ti-layers on all sides have been developed (activity 0.5 - 2 GBq). Quality checks with plastic scintillators have been developed and are correlated to absolute dose measurements performed with TLDs (1x1 mm2; 40 mg/cm2). TLD-handling and calibration for beta-dosimetry are described. Additional measurements for depth-dose and dose distribution around the tubes were done with GAFCHROMIC- films and compared to Monte-Carlo calculations with the MCNP4-code, yielding a half-value depth of 0.8 mm from the tube-surface. Manufacturing and delivery of the sources including leakage tests has been standardized, treatments (irradiation times <5min; irradiation length <30mm) are planned to start in spring `98. (authors) 1 refs., 5 figs.
Authors:
Scmidt, W F.O.; Hawliczek, R; [1]  Mueck, K; [2]  Lehmann, D; [3]  Pichler, L [4] 
  1. Inst of Radiooncology IRO, Donauspital, Vienna (Austria)
  2. Austrian Research Centre, Siebersdorf ARCS (Austria)
  3. Inst of Radiotherapy, Univ. Dresden (Germany)
  4. Ludwig Boltzmann Institute for Endovascular Radiotherapy, Donauspital, Vienna (Austria)
Publication Date:
Dec 31, 1998
Product Type:
Conference
Report Number:
INIS-CY-0004; CONF-9806202-
Reference Number:
SCA: 550604; PA: AIX-30:017097; EDB-99:043451; SN: 99002082140
Resource Relation:
Conference: 8. Mediterranean conference on medical and biological engineering and computing (Medicon `98), Lemesos (Cyprus), 14-17 Jun 1998; Other Information: DN: 1 refs., 5 figs. The proceedings of this meeting are available on CD-ROM in PDF, at the price of US$50, from the Cyprus Association of Medical Physics and Biomedical Engineering, P.O.Box 24039, 1700 Nicosia, Cyprus. The CD-ROM makes use of Abode Acrobat PDF (Portable Document Format). To read the files the Abode Acrobat reader has to be installed on your system. A copy is provided with the CD-ROM; PBD: 1998; Related Information: Is Part Of Proceedings of the 8. Mediterranean Conference on Medical and Biological Engineering and Computing (Medicon `98); Christofides, Stelios; Pattichis, Constantinos; Schizas, Christos; Keravnou-Papailiou, Elpida; Kaplanis, Prodromos; Spyros, Spyrou; Christodoulides, George; Theodoulou, Yiannis [eds.]; PB: 220 p.
Subject:
55 BIOLOGY AND MEDICINE, BASIC STUDIES; BETA DOSIMETRY; DEPTH DOSE DISTRIBUTIONS; MONTE CARLO METHOD; RADIATION DOSE DISTRIBUTIONS; RADIATION SOURCE IMPLANTS; YTTRIUM 90
OSTI ID:
331904
Research Organizations:
Cyprus Association of Medical Physics and Biomedical Engineering (CAMPBE), Nicosia (Cyprus); The Department of Computer Science, University of Cyprus, Nicosia (Cyprus)
Country of Origin:
Cyprus
Language:
English
Other Identifying Numbers:
Other: ON: DE99619912; ISBN 9963-607-14-4; TRN: CY9900020017097
Availability:
INIS; OSTI as DE99619912
Submitting Site:
INIS
Size:
pp. 131
Announcement Date:

Citation Formats

Scmidt, W F.O., Hawliczek, R, Mueck, K, Lehmann, D, and Pichler, L. Endovascular Irradiations with beta sources. Cyprus: N. p., 1998. Web.
Scmidt, W F.O., Hawliczek, R, Mueck, K, Lehmann, D, & Pichler, L. Endovascular Irradiations with beta sources. Cyprus.
Scmidt, W F.O., Hawliczek, R, Mueck, K, Lehmann, D, and Pichler, L. 1998. "Endovascular Irradiations with beta sources." Cyprus.
@misc{etde_331904,
title = {Endovascular Irradiations with beta sources}
author = {Scmidt, W F.O., Hawliczek, R, Mueck, K, Lehmann, D, and Pichler, L}
abstractNote = {For treatment of restenoses tubes (inner/outer diameter 1 and 2 mm; length 3 or 5 mm) with Y-90 foils, shielded by Ti-layers on all sides have been developed (activity 0.5 - 2 GBq). Quality checks with plastic scintillators have been developed and are correlated to absolute dose measurements performed with TLDs (1x1 mm2; 40 mg/cm2). TLD-handling and calibration for beta-dosimetry are described. Additional measurements for depth-dose and dose distribution around the tubes were done with GAFCHROMIC- films and compared to Monte-Carlo calculations with the MCNP4-code, yielding a half-value depth of 0.8 mm from the tube-surface. Manufacturing and delivery of the sources including leakage tests has been standardized, treatments (irradiation times <5min; irradiation length <30mm) are planned to start in spring `98. (authors) 1 refs., 5 figs.}
place = {Cyprus}
year = {1998}
month = {Dec}
}