Abstract
The Fricke dosemeter has been used for absolute determination of the radiation dose between 40 and 350 Gy for X-rays, Gamma rays and fast electrons energies. The great inconvenient of that dosimetric system is the impossibility of using for measurements of exposure or low dose rates. A Fricke dosimetry system was implanted at the Nuclear Sciences Regional Center (CRCN), in Recife, Pernambuco state, Brazil, for the determination of radiation doses for spectrometry and mass dosimetry with plasma source. The statistical analysis of the obtained results have shown that using the spectrometric method that system cna be used for maintenance of factors of CRCN standard calibrations and the dose assessment thru mail system of radiotherapy centers of the country.
Pedrosa, G A.C. ,;
[1]
Dantas, C.C. , e-mail: ccd@ufpe.br;
[2]
Campos, C.A. , e-mail: auster@elogica.com.br
[3]
- Instituto Nacional de Metrologia (INMETRO), Rio de Janeiro, RJ (Brazil)
- Universidade Federal de Pernambuco (UFPE), Recife, PE (Brazil). Dept. de Energia Nuclear
- Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE/CNEN-PE), Recife, PE (Brazil)
Citation Formats
Pedrosa, G A.C. ,, Dantas, C.C. , e-mail: ccd@ufpe.br, and Campos, C.A. , e-mail: auster@elogica.com.br.
Measurement analysis using the Fricke dosemeter; Analise de medicoes com o dosimetro Fricke.
Brazil: N. p.,
2003.
Web.
Pedrosa, G A.C. ,, Dantas, C.C. , e-mail: ccd@ufpe.br, & Campos, C.A. , e-mail: auster@elogica.com.br.
Measurement analysis using the Fricke dosemeter; Analise de medicoes com o dosimetro Fricke.
Brazil.
Pedrosa, G A.C. ,, Dantas, C.C. , e-mail: ccd@ufpe.br, and Campos, C.A. , e-mail: auster@elogica.com.br.
2003.
"Measurement analysis using the Fricke dosemeter; Analise de medicoes com o dosimetro Fricke."
Brazil.
@misc{etde_21203100,
title = {Measurement analysis using the Fricke dosemeter; Analise de medicoes com o dosimetro Fricke}
author = {Pedrosa, G A.C. ,, Dantas, C.C. , e-mail: ccd@ufpe.br, and Campos, C.A. , e-mail: auster@elogica.com.br}
abstractNote = {The Fricke dosemeter has been used for absolute determination of the radiation dose between 40 and 350 Gy for X-rays, Gamma rays and fast electrons energies. The great inconvenient of that dosimetric system is the impossibility of using for measurements of exposure or low dose rates. A Fricke dosimetry system was implanted at the Nuclear Sciences Regional Center (CRCN), in Recife, Pernambuco state, Brazil, for the determination of radiation doses for spectrometry and mass dosimetry with plasma source. The statistical analysis of the obtained results have shown that using the spectrometric method that system cna be used for maintenance of factors of CRCN standard calibrations and the dose assessment thru mail system of radiotherapy centers of the country.}
place = {Brazil}
year = {2003}
month = {Jul}
}
title = {Measurement analysis using the Fricke dosemeter; Analise de medicoes com o dosimetro Fricke}
author = {Pedrosa, G A.C. ,, Dantas, C.C. , e-mail: ccd@ufpe.br, and Campos, C.A. , e-mail: auster@elogica.com.br}
abstractNote = {The Fricke dosemeter has been used for absolute determination of the radiation dose between 40 and 350 Gy for X-rays, Gamma rays and fast electrons energies. The great inconvenient of that dosimetric system is the impossibility of using for measurements of exposure or low dose rates. A Fricke dosimetry system was implanted at the Nuclear Sciences Regional Center (CRCN), in Recife, Pernambuco state, Brazil, for the determination of radiation doses for spectrometry and mass dosimetry with plasma source. The statistical analysis of the obtained results have shown that using the spectrometric method that system cna be used for maintenance of factors of CRCN standard calibrations and the dose assessment thru mail system of radiotherapy centers of the country.}
place = {Brazil}
year = {2003}
month = {Jul}
}