Calculated Absolute Detection Efficiencies of Cylindrical Nal (Tl) Scintillation Crystals for Aqueous Spherical Sources

Abstract

Calculated values of the absolute total detection efficiencies of cylindrical scintillation crystals viewing spherical sources of various sizes are presented. The calculation is carried out for 2 x 2 inch and 3 x 3 inch Nal(Tl) crystals and for sources which have the radii 1/4, 1/2, 3/4 and 1 times the crystal radius. Source-detector distances of 5-20 cm and gamma energies in the range 0.1 - 5 MeV are considered. The correction factor for absorption in the sample container wall and in the detector housing is derived and calculated for a practical case.
Publication Date:
Aug 15, 1968
Product Type:
Technical Report
Report Number:
AE-310
Resource Relation:
Other Information: 7 refs., 2 figs., 4 tabs.
Subject:
73 NUCLEAR PHYSICS AND RADIATION PHYSICS; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; EFFICIENCY; ENERGY DEPENDENCE; NAI DETECTORS; SPHERES; GAMMA DETECTION; DISTANCE; RADIATION SOURCES
OSTI ID:
20956286
Research Organizations:
AB Atomenergi, Nykoeping (Sweden)
Country of Origin:
Sweden
Language:
English
Other Identifying Numbers:
TRN: SE0708720
Availability:
Commercial reproduction prohibited; OSTI as DE20956286
Submitting Site:
SWDN
Size:
22 pages
Announcement Date:
Dec 29, 2007

Citation Formats

Strindehag, O, and Tollander, B. Calculated Absolute Detection Efficiencies of Cylindrical Nal (Tl) Scintillation Crystals for Aqueous Spherical Sources. Sweden: N. p., 1968. Web.
Strindehag, O, & Tollander, B. Calculated Absolute Detection Efficiencies of Cylindrical Nal (Tl) Scintillation Crystals for Aqueous Spherical Sources. Sweden.
Strindehag, O, and Tollander, B. 1968. "Calculated Absolute Detection Efficiencies of Cylindrical Nal (Tl) Scintillation Crystals for Aqueous Spherical Sources." Sweden.
@misc{etde_20956286,
title = {Calculated Absolute Detection Efficiencies of Cylindrical Nal (Tl) Scintillation Crystals for Aqueous Spherical Sources}
author = {Strindehag, O, and Tollander, B}
abstractNote = {Calculated values of the absolute total detection efficiencies of cylindrical scintillation crystals viewing spherical sources of various sizes are presented. The calculation is carried out for 2 x 2 inch and 3 x 3 inch Nal(Tl) crystals and for sources which have the radii 1/4, 1/2, 3/4 and 1 times the crystal radius. Source-detector distances of 5-20 cm and gamma energies in the range 0.1 - 5 MeV are considered. The correction factor for absorption in the sample container wall and in the detector housing is derived and calculated for a practical case.}
place = {Sweden}
year = {1968}
month = {Aug}
}