Study of the performance of a 4 {pi} {gamma} ionisation chamber; Etude des performances d'une chambre d'ionisation 4 {pi} {gamma}

Abstract

Description of a well-type {gamma} chamber for routine measurements of sources between 1 {mu}curie and 1 curie. Approximate calculation of some geometrical parameters by which the zone of constant sensitivity may be increased. Calculation of the sensitivity of such a chamber and comparison with the experimental results. (author) [French] Description d'une chambre a puits {gamma}, destinee aux mesures de routine de sources entre 1 {mu}curie et 1 curie. Calcul approche de quelques parametres geometriques permettant d'accroitre la zone d'egale sensibilite. Calcul de la sensibilite d'une telle chambre et confrontation avec les resultats experimentaux. (auteur)
Authors:
Engelmann, J J [1] 
  1. Commissariat a l'Energie Atomique, Saclay (France).Centre d'Etudes Nucleaires
Publication Date:
Jul 01, 1960
Product Type:
Technical Report
Report Number:
CEA-R-1539
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ANNULAR SPACE; ARGON; CALIBRATION; CURRENTS; EXTRAPOLATION; GAMMA DETECTION; GAMMA SOURCES; IONIZATION CHAMBERS; KEV RANGE 100-1000; MEV RANGE 01-10; PERFORMANCE TESTING; PRESSURE DEPENDENCE; SAMPLE HOLDERS; SENSITIVITY; SPACE DEPENDENCE; SPECIFICATIONS
OSTI ID:
20951827
Research Organizations:
CEA Saclay, 91 - Gif-sur-Yvette (France)
Country of Origin:
France
Language:
French
Other Identifying Numbers:
TRN: FR07R1539102053
Availability:
Available from INIS in electronic form
Submitting Site:
FRN
Size:
27 pages
Announcement Date:
Dec 21, 2007

Citation Formats

Engelmann, J J. Study of the performance of a 4 {pi} {gamma} ionisation chamber; Etude des performances d'une chambre d'ionisation 4 {pi} {gamma}. France: N. p., 1960. Web.
Engelmann, J J. Study of the performance of a 4 {pi} {gamma} ionisation chamber; Etude des performances d'une chambre d'ionisation 4 {pi} {gamma}. France.
Engelmann, J J. 1960. "Study of the performance of a 4 {pi} {gamma} ionisation chamber; Etude des performances d'une chambre d'ionisation 4 {pi} {gamma}." France.
@misc{etde_20951827,
title = {Study of the performance of a 4 {pi} {gamma} ionisation chamber; Etude des performances d'une chambre d'ionisation 4 {pi} {gamma}}
author = {Engelmann, J J}
abstractNote = {Description of a well-type {gamma} chamber for routine measurements of sources between 1 {mu}curie and 1 curie. Approximate calculation of some geometrical parameters by which the zone of constant sensitivity may be increased. Calculation of the sensitivity of such a chamber and comparison with the experimental results. (author) [French] Description d'une chambre a puits {gamma}, destinee aux mesures de routine de sources entre 1 {mu}curie et 1 curie. Calcul approche de quelques parametres geometriques permettant d'accroitre la zone d'egale sensibilite. Calcul de la sensibilite d'une telle chambre et confrontation avec les resultats experimentaux. (auteur)}
place = {France}
year = {1960}
month = {Jul}
}