Abstract
A method for as certain the activity by alpha spectroscopy with semiconductor detectors, of a solution of 232U is presented. It consists of the comparison with a 233U solution activity previously measured in a gridded ionization chamber of 2{pi} geometry. The total measurement uncertainty is about + - 0,02. (Author) 9 refs.
Citation Formats
Galan Valera, M P, and Acena Berrenechea, M L.
Calibration of 232U solution; Calibracion de una disolucion de 232U.
Spain: N. p.,
1988.
Web.
Galan Valera, M P, & Acena Berrenechea, M L.
Calibration of 232U solution; Calibracion de una disolucion de 232U.
Spain.
Galan Valera, M P, and Acena Berrenechea, M L.
1988.
"Calibration of 232U solution; Calibracion de una disolucion de 232U."
Spain.
@misc{etde_20892652,
title = {Calibration of 232U solution; Calibracion de una disolucion de 232U}
author = {Galan Valera, M P, and Acena Berrenechea, M L}
abstractNote = {A method for as certain the activity by alpha spectroscopy with semiconductor detectors, of a solution of 232U is presented. It consists of the comparison with a 233U solution activity previously measured in a gridded ionization chamber of 2{pi} geometry. The total measurement uncertainty is about + - 0,02. (Author) 9 refs.}
place = {Spain}
year = {1988}
month = {Jul}
}
title = {Calibration of 232U solution; Calibracion de una disolucion de 232U}
author = {Galan Valera, M P, and Acena Berrenechea, M L}
abstractNote = {A method for as certain the activity by alpha spectroscopy with semiconductor detectors, of a solution of 232U is presented. It consists of the comparison with a 233U solution activity previously measured in a gridded ionization chamber of 2{pi} geometry. The total measurement uncertainty is about + - 0,02. (Author) 9 refs.}
place = {Spain}
year = {1988}
month = {Jul}
}