Electron multiplier for the measurement of an ion current on a mass spectrometer; Multiplicateur d'electrons pour la mesure de courant d'ions sur un spectrometre de masse

Abstract

The apparatus described is designed to measure weak ion currents received at the collector of a mass spectrometer. The report describes successively the study of electron paths in the multiplier by the method of analogy, using rubber membranes, and the practical details of construction of the apparatus. The variation with surface treatment of the secondary emission coefficient of the alloy CuBe containing 2 per cent Be, which makes up the dynodes, and the influence of the voltage on the gain per stage, are discussed. Results of tests regarding: the influence of the ion mass on the gain, the background of the instrument and the energy distribution of the impulses coming out on a high gain multiplier (q.q. 10{sup 7}) are given. Finally the performances of the multiplier are reported. 1- For a low gain (10{sup 4}), precision and reproducibility comparable to the electrometer valve, sensitivity 100 times greater, currents capable of detection 10{sup -17} Ampere. 2- For a high gain (10{sup 7}) and measurement by impulse counting, currents capable of detection 10{sup -19} Ampere. Mounting difficult to use on a mass spectrometer. (author) [French] L'appareil decrit est destine a la mesure des faibles courants d'ions re s au collecteur d'un  More>>
Authors:
Lohez, P; Nief, G [1] 
  1. Commissariat a l'Energie Atomique, Saclay (France). Centre d'Etudes Nucleaires
Publication Date:
Jul 01, 1957
Product Type:
Technical Report
Report Number:
CEA-R-692
Resource Relation:
Other Information: 3 refs
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ACCURACY; BACKGROUND NOISE; BERYLLIUM ALLOYS; COPPER BASE ALLOYS; COUNTING TECHNIQUES; DYNODES; ELECTRON MULTIPLIERS; ELECTROPOLISHING; GAIN; ION BEAMS; MASS SPECTROMETERS; MECHANICAL POLISHING; PULSE TECHNIQUES; SECONDARY EMISSION; SENSITIVITY; SURFACE TREATMENTS
OSTI ID:
20877813
Research Organizations:
CEA Saclay, 91 - Gif-sur-Yvette (France)
Country of Origin:
France
Language:
French
Other Identifying Numbers:
TRN: FR07R0692045526
Availability:
Available from INIS in electronic form
Submitting Site:
FRN
Size:
28 pages
Announcement Date:
Jun 15, 2007

Citation Formats

Lohez, P, and Nief, G. Electron multiplier for the measurement of an ion current on a mass spectrometer; Multiplicateur d'electrons pour la mesure de courant d'ions sur un spectrometre de masse. France: N. p., 1957. Web.
Lohez, P, & Nief, G. Electron multiplier for the measurement of an ion current on a mass spectrometer; Multiplicateur d'electrons pour la mesure de courant d'ions sur un spectrometre de masse. France.
Lohez, P, and Nief, G. 1957. "Electron multiplier for the measurement of an ion current on a mass spectrometer; Multiplicateur d'electrons pour la mesure de courant d'ions sur un spectrometre de masse." France.
@misc{etde_20877813,
title = {Electron multiplier for the measurement of an ion current on a mass spectrometer; Multiplicateur d'electrons pour la mesure de courant d'ions sur un spectrometre de masse}
author = {Lohez, P, and Nief, G}
abstractNote = {The apparatus described is designed to measure weak ion currents received at the collector of a mass spectrometer. The report describes successively the study of electron paths in the multiplier by the method of analogy, using rubber membranes, and the practical details of construction of the apparatus. The variation with surface treatment of the secondary emission coefficient of the alloy CuBe containing 2 per cent Be, which makes up the dynodes, and the influence of the voltage on the gain per stage, are discussed. Results of tests regarding: the influence of the ion mass on the gain, the background of the instrument and the energy distribution of the impulses coming out on a high gain multiplier (q.q. 10{sup 7}) are given. Finally the performances of the multiplier are reported. 1- For a low gain (10{sup 4}), precision and reproducibility comparable to the electrometer valve, sensitivity 100 times greater, currents capable of detection 10{sup -17} Ampere. 2- For a high gain (10{sup 7}) and measurement by impulse counting, currents capable of detection 10{sup -19} Ampere. Mounting difficult to use on a mass spectrometer. (author) [French] L'appareil decrit est destine a la mesure des faibles courants d'ions re s au collecteur d'un spectrometre de masse. Le rapport decrit successivement l'etude des trajectoires des electrons dans le multiplicateur, par la methode analogique de la menbrane en caoutchouc, et la realisation pratique de l'appareil. La variation du coefficient d'emission secondaire de l'alliage CuBe a 2 pour cent de Be, constituant les dynodes suivant le traitement des surfaces, et l'influence de la tension sur le gain par etage sont discutees. Des resultats d'essais concernant: l'influence de la masse des ions sur le gain, le bruit de fond de l'appareil et la repartition en energie des impulsions de sortie sur un multiplicateur a gain eleve (q.q. 10{sup 7}) sont donnes. Enfin, sont rapportees les performances du multiplicateur. 1- pour un gain faible (10{sup 4}) precision et reproductibilite comparables a la lampe electrometre, sensibilite 100 fois plus grande courants decelables 1O{sup -17} Ampere. 2- pour un gain eleve (10{sup 7} ) et mesure par comptage d'impulsions, courants decelables 10{sup -19} Ampere. Montage difficilement employable sur un spectrometre de masse. (auteur)}
place = {France}
year = {1957}
month = {Jul}
}