You need JavaScript to view this

Isothermal calorimeter for reactor radiation dosimetry

Abstract

An isothermal calorimeter with thermistors for measuring absorbed dose rates from 10{sup 4}-5-6.10{sup 5} rad/h in reactor experimental holes has been designed. A kinetics method for determining the equilibrium temperature difference has been developed, and its application in isothermal calorimetry proved. The expected accuracy in measurements within {+-} 2-5% has been proved by measurements carried out in the reactor. Some data obtained by measurements in the reactor RA are presented (author)
Authors:
Radak, B; Markovic, V [1] 
  1. Institute of Nuclear Sciences Boris Kidric, Odeljenje za radijacionu hemiju, Vinca, Beograd (Serbia and Montenegro)
Publication Date:
Dec 15, 1961
Product Type:
Technical Report
Report Number:
INIS-CS-0281; IZ-051-0061-1961
Resource Relation:
Other Information: 3 tabs, 6 figs, 8 refs; 14 tabs, 30 figs, 15 refs; Related Information: In: Calorimeter measurements of absorbed doses at the heavy water enriched uranium reactor, Kalorimetrijska merenja apsorbovanih doza u reaktoru na tesku vodu i obogaceni uran, by Markovic, V. [Institute of Nuclear Sciences Boris Kidric, Odeljenje za radijacionu hemiju, Vinca, Beograd (Serbia and Montenegro)], 101 pages.
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ACCURACY; CALORIMETRIC DOSEMETERS; EXPERIMENTAL CHANNELS; R-A REACTOR; THERMAL EQUILIBRIUM; THERMISTORS
OSTI ID:
20871694
Research Organizations:
Savezna komisija za nuklearnu energiju (Yugoslavia)
Country of Origin:
Serbia and Montenegro
Language:
English
Other Identifying Numbers:
TRN: CS06RA745039531
Availability:
Available from INIS in electronic form; Also available from the Institute of nuclear sciences Vinca
Submitting Site:
INIS
Size:
page(s) 19
Announcement Date:
Jun 11, 2007

Citation Formats

Radak, B, and Markovic, V. Isothermal calorimeter for reactor radiation dosimetry. Serbia and Montenegro: N. p., 1961. Web.
Radak, B, & Markovic, V. Isothermal calorimeter for reactor radiation dosimetry. Serbia and Montenegro.
Radak, B, and Markovic, V. 1961. "Isothermal calorimeter for reactor radiation dosimetry." Serbia and Montenegro.
@misc{etde_20871694,
title = {Isothermal calorimeter for reactor radiation dosimetry}
author = {Radak, B, and Markovic, V}
abstractNote = {An isothermal calorimeter with thermistors for measuring absorbed dose rates from 10{sup 4}-5-6.10{sup 5} rad/h in reactor experimental holes has been designed. A kinetics method for determining the equilibrium temperature difference has been developed, and its application in isothermal calorimetry proved. The expected accuracy in measurements within {+-} 2-5% has been proved by measurements carried out in the reactor. Some data obtained by measurements in the reactor RA are presented (author)}
place = {Serbia and Montenegro}
year = {1961}
month = {Dec}
}