Abstract

The behaviour of a proportional BF{sub 3}LCT 14 NE 31 counter with aluminium cathode has been studied after irradiation from 10{sup 5} to 10{sup 6} n.cm{sup -2} in the swimming-pool reactor Triton. The pulse spectrum was only slightly modified by two successive irradiations at 10{sup 15} n.cm{sup -2}, and 1 hour after the end of the irradiation the decrease in sensitivity reached 10 to 20 per cent. One hour after irradiation at 10{sup 16} n.cm{sup -2} this decrease was 20 per cent but the pulse spectrum was affected. The advantages of an aluminium cathode with respect to a standard counter using a copper cathode and irradiated under the same conditions are given. Beforehand, it had been established that the limit for use of an aluminium-cathode counter in the presence of {gamma} radiation is about 200 R.h{sup -1}. (author) [French] Le comportement d'un compteur proportionnel a BF-3LCT 14 NE 31, a cathode d'aluminium, a ete etudie apres des irradiations de 10{sup 5} a 10{sup 6} n.cm{sup -2} dans la pile-piscine TRITON. Deux irradiations successives de 10{sup 15} n.cm{sup -2} ont peu modifie le spectre des impulsions et 1 heure apres la fin de l'exposition la perte de sensibilite etait de 10  More>>
Authors:
Verdant, R [1] 
  1. Commissariat a l'Energie Atomique, Saclay (France). Centre d'Etudes Nucleaires
Publication Date:
Jul 01, 1967
Product Type:
Technical Report
Report Number:
CEA-R-3343
Resource Relation:
Other Information: 6 refs
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ALUMINIUM; BF3 COUNTERS; CATHODES; COMPARATIVE EVALUATIONS; GAMMA RADIATION; IRRADIATION; PERFORMANCE TESTING; SENSITIVITY
OSTI ID:
20726979
Research Organizations:
CEA Saclay, 91 - Gif-sur-Yvette (France)
Country of Origin:
France
Language:
French
Other Identifying Numbers:
TRN: FR06R3343038424
Availability:
Available from INIS in electronic form
Submitting Site:
FRN
Size:
36 pages
Announcement Date:
May 24, 2006

Citation Formats

Verdant, R. Behaviour of BF{sub 3} counters after in-pile irradiation; Comportement de compteurs a BF{sub 3} apres irradiation dans un reacteur. France: N. p., 1967. Web.
Verdant, R. Behaviour of BF{sub 3} counters after in-pile irradiation; Comportement de compteurs a BF{sub 3} apres irradiation dans un reacteur. France.
Verdant, R. 1967. "Behaviour of BF{sub 3} counters after in-pile irradiation; Comportement de compteurs a BF{sub 3} apres irradiation dans un reacteur." France.
@misc{etde_20726979,
title = {Behaviour of BF{sub 3} counters after in-pile irradiation; Comportement de compteurs a BF{sub 3} apres irradiation dans un reacteur}
author = {Verdant, R}
abstractNote = {The behaviour of a proportional BF{sub 3}LCT 14 NE 31 counter with aluminium cathode has been studied after irradiation from 10{sup 5} to 10{sup 6} n.cm{sup -2} in the swimming-pool reactor Triton. The pulse spectrum was only slightly modified by two successive irradiations at 10{sup 15} n.cm{sup -2}, and 1 hour after the end of the irradiation the decrease in sensitivity reached 10 to 20 per cent. One hour after irradiation at 10{sup 16} n.cm{sup -2} this decrease was 20 per cent but the pulse spectrum was affected. The advantages of an aluminium cathode with respect to a standard counter using a copper cathode and irradiated under the same conditions are given. Beforehand, it had been established that the limit for use of an aluminium-cathode counter in the presence of {gamma} radiation is about 200 R.h{sup -1}. (author) [French] Le comportement d'un compteur proportionnel a BF-3LCT 14 NE 31, a cathode d'aluminium, a ete etudie apres des irradiations de 10{sup 5} a 10{sup 6} n.cm{sup -2} dans la pile-piscine TRITON. Deux irradiations successives de 10{sup 15} n.cm{sup -2} ont peu modifie le spectre des impulsions et 1 heure apres la fin de l'exposition la perte de sensibilite etait de 10 a 20 pour cent. Une heure apres une irradiation de 10{sup 16} n.cm{sup -2}, cette perte etait de 20 pour cent mais le spectre d'impulsions etait perturbe. L'avantage d'une cathode en aluminium est mis en evidence par comparaison avec un compteur standard a cathode de cuivre irradie dans les memes conditions. En preliminaire, on avait etabli que la limite d'utilisation d'un compteur a cathode d'aluminium en presence de rayonnements {gamma} etait de l'ordre de 200 R.h{sup -1}. (auteur)}
place = {France}
year = {1967}
month = {Jul}
}