Detection of hard {beta} radiation in water using the Cerenkov effect; Detection des {beta} durs dans l'eau par effet Cerenkov

Abstract

The light emitted from a liquid during the passage of a particle of sufficiently high energy can be used to determine the contamination of water by {beta}-emitters. The detector based on this principle requires no special treatment of the water. Particular attention has been paid to the optical yield of the detector. However the excellent sensitivity to which this principle leads can only be attained if the background noise is considerably reduced. The electronics associated with detector make it possible to detect a contamination of 8 x 10{sup -8} {mu}c/cm{sup 3} of strontium-90 in water in about ten minutes. The apparatus presents the double advantage of having a high sensitivity and of being easy to use. Its characteristics make it possible for it to be used for the continuous control of water. (author) [French] La lumiere emise au sein d'un liquide au passage d'une particule suffisamment energetique peut permettre de determiner la contamination d'une eau en emetteurs {beta}. Le detecteur realise sur ce principe ne necessite aucun traitement special de l'eau. Un soin tout particulier a ete apporte a l'etude du rendement optique du detecteur. Cependant l'excellente sensibilite attachee au principe ne peut etre utilisee que si l'on reduit considerablement  More>>
Authors:
Turck, B [1] 
  1. Commissariat a l'Energie Atomique, Saclay (France). Centre d'Etudes Nucleaires
Publication Date:
Jan 01, 1965
Product Type:
Thesis/Dissertation
Report Number:
CEA-R-2915
Resource Relation:
Other Information: TH: These ES sciences physiques; 28 refs; PBD: Jan 1965
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ABSORPTION SPECTRA; BACKGROUND NOISE; BETA DETECTION; CALIBRATION; CHERENKOV COUNTERS; COINCIDENCE CIRCUITS; COINCIDENCE METHODS; COMPTON EFFECT; EFFICIENCY; GAMMA DETECTION; OPTICAL REFLECTION; PHOTOCATHODES; PHOTOELECTRON COUNTING; PHOTOMULTIPLIERS; PHOTONS; PROBABILISTIC ESTIMATION; RESOLUTION; SENSITIVITY; SIGNAL-TO-NOISE RATIO; SPECTRAL RESPONSE; STRONTIUM 90; TRAJECTORIES; WATER; YTTRIUM 90
OSTI ID:
20598738
Research Organizations:
CEA Saclay, 91 - Gif-sur-Yvette (France); Faculte des Sciences de l'Universite de Paris, Centre d'Orsay, 91 (France)
Country of Origin:
France
Language:
French
Other Identifying Numbers:
TRN: FR05R2915045250
Availability:
Available from INIS in electronic form
Submitting Site:
FRN
Size:
69 pages
Announcement Date:
Jun 12, 2005

Citation Formats

Turck, B. Detection of hard {beta} radiation in water using the Cerenkov effect; Detection des {beta} durs dans l'eau par effet Cerenkov. France: N. p., 1965. Web.
Turck, B. Detection of hard {beta} radiation in water using the Cerenkov effect; Detection des {beta} durs dans l'eau par effet Cerenkov. France.
Turck, B. 1965. "Detection of hard {beta} radiation in water using the Cerenkov effect; Detection des {beta} durs dans l'eau par effet Cerenkov." France.
@misc{etde_20598738,
title = {Detection of hard {beta} radiation in water using the Cerenkov effect; Detection des {beta} durs dans l'eau par effet Cerenkov}
author = {Turck, B}
abstractNote = {The light emitted from a liquid during the passage of a particle of sufficiently high energy can be used to determine the contamination of water by {beta}-emitters. The detector based on this principle requires no special treatment of the water. Particular attention has been paid to the optical yield of the detector. However the excellent sensitivity to which this principle leads can only be attained if the background noise is considerably reduced. The electronics associated with detector make it possible to detect a contamination of 8 x 10{sup -8} {mu}c/cm{sup 3} of strontium-90 in water in about ten minutes. The apparatus presents the double advantage of having a high sensitivity and of being easy to use. Its characteristics make it possible for it to be used for the continuous control of water. (author) [French] La lumiere emise au sein d'un liquide au passage d'une particule suffisamment energetique peut permettre de determiner la contamination d'une eau en emetteurs {beta}. Le detecteur realise sur ce principe ne necessite aucun traitement special de l'eau. Un soin tout particulier a ete apporte a l'etude du rendement optique du detecteur. Cependant l'excellente sensibilite attachee au principe ne peut etre utilisee que si l'on reduit considerablement le bruit de fond. L'electronique associee au detecteur permet de reperer l'existence d'une contamination de 8 x 10{sup -8} {mu}c/cm{sup 3} de strontium-90 dans l'eau en un temps d'une dizaine de minutes. L'appareil presente l'avantage d'allier une grande sensibilite a la facilite d'emploi. Ses performances autorisent son utilisation au controle continu des eaux. (auteur)}
place = {France}
year = {1965}
month = {Jan}
}