You need JavaScript to view this

Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear

Abstract

With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some {mu} Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some {mu} Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm{sup  More>>
Authors:
Cortes P, A; Garcia D, O C; Ortiz P, I; Becerril V, A [1] 
  1. Instituto nacional de Investigaciones Nucleares, Laboratorio de Patrones Radiactivos, Departamento de Metrologia, A.P. 18-1027. C.P. 11801 Mexico D.F. (Mexico)
Publication Date:
Jul 01, 2000
Product Type:
Conference
Report Number:
INIS-MX-696
Reference Number:
EDB-01:027315
Resource Relation:
Conference: 10. ININ-SUTIN Technical and Scientific Congress, Salazar, Estado de Mexico (Mexico), 29-30 Nov 2000; Other Information: PBD: 2000
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; ACTIVITY METERS; BARIUM 133; CALIBRATION; CESIUM 137; COBALT 60; DIAGRAMS; EXPERIMENTAL DATA; FRENCH ORGANIZATIONS; GALLIUM 67; HALF-LIFE; INDIUM 111; IODINE 131; MEXICO; MOLYBDENUM 99; NUCLEAR MEDICINE; SAMARIUM 153; TECHNETIUM 99; THALLIUM 201
OSTI ID:
20143516
Research Organizations:
Instituto Nacional de Investigaciones Nucleares. Sindicato Unico de Trabajadores de la Industria Nuclear (Mexico)
Country of Origin:
Mexico
Language:
Spanish
Other Identifying Numbers:
TRN: MX0100117010592
Availability:
Available from INIS in electronic form
Submitting Site:
MXN
Size:
5 pages
Announcement Date:
Mar 27, 2001

Citation Formats

Cortes P, A, Garcia D, O C, Ortiz P, I, and Becerril V, A. Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear. Mexico: N. p., 2000. Web.
Cortes P, A, Garcia D, O C, Ortiz P, I, & Becerril V, A. Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear. Mexico.
Cortes P, A, Garcia D, O C, Ortiz P, I, and Becerril V, A. 2000. "Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear." Mexico.
@misc{etde_20143516,
title = {Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear}
author = {Cortes P, A, Garcia D, O C, Ortiz P, I, and Becerril V, A}
abstractNote = {With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some {mu} Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some {mu} Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm{sup 3} contained in polyethylene small bottles of 25 cm{sup 3} elaborated in the LPR and whose activity is measured with respect to the National Patron ININ-PNM-2. The sources of mid-large life are ampoules with 5 cm{sup 3} of radioactive solution, calibrated in activity by the primary laboratory, LNHB (formerly LMRI) of France. (Author)}
place = {Mexico}
year = {2000}
month = {Jul}
}