You need JavaScript to view this

The approbation of rotating scanning irradiation method at therapeutic proton beam of the Laboratory of Nuclear Problems, JINR; Aprobatsiya metoda rotatsionnogo skaniruyushchego oblucheniya na meditsinskom protonnom puchke Laboratorii yadernykh problem, OIYaI

Abstract

The method of rotating scanning irradiation with a proton beam for deeply located tumours is described. The Bragg peak and the tumour overlapping are controlled by a computer on the basis of information from miniature silicon detector inserted in the cavity inside the patient body near the tumour or from calculations from X-ray tomographic images, measured immediately before irradiation runs. In phantom tests good agreement between calculations from tomographic images and measurements with silicon detector has been obtained. The method may be used in clinical research. (author). 13 refs.; 12 figs.
Authors:
Abazov, V M; Budyashov, Yu G; Zorin, V P; Mitsyn, G V; Molokanov, A G; Savchenko, O V; [1]  Astrakhan, B V [2] 
  1. Joint Inst. for Nuclear Research, Dubna (Russian Federation). Lab. of Nuclear Problems
  2. Akademiya Meditsinskikh Nauk SSSR, Moscow (Russian Federation). Inst. Onkologii
Publication Date:
Dec 31, 1993
Product Type:
Technical Report
Report Number:
JINR-18-93-351
Reference Number:
SCA: 550603; PA: AIX-25:031286; EDB-94:070967; NTS-94:020603; SN: 94001192345
Resource Relation:
Other Information: PBD: 1993
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; RADIOTHERAPY; NEOPLASMS; PROTON BEAMS; BEAM SCANNERS; BRAGG CURVE; COMPUTERIZED TOMOGRAPHY; DEPTH DOSE DISTRIBUTIONS; EXTERNAL IRRADIATION; PHANTOMS; ROTATION; 550603; EXTERNAL RADIATION IN THERAPY
OSTI ID:
10144755
Research Organizations:
Joint Inst. for Nuclear Research, Dubna (Russian Federation). Lab. of Nuclear Problems
Country of Origin:
JINR
Language:
Russian
Other Identifying Numbers:
Other: ON: DE94622426; TRN: XJ9406297031286
Availability:
OSTI; NTIS (US Sales Only); INIS
Submitting Site:
INIS
Size:
18 p.
Announcement Date:
Jul 05, 2005

Citation Formats

Abazov, V M, Budyashov, Yu G, Zorin, V P, Mitsyn, G V, Molokanov, A G, Savchenko, O V, and Astrakhan, B V. The approbation of rotating scanning irradiation method at therapeutic proton beam of the Laboratory of Nuclear Problems, JINR; Aprobatsiya metoda rotatsionnogo skaniruyushchego oblucheniya na meditsinskom protonnom puchke Laboratorii yadernykh problem, OIYaI. JINR: N. p., 1993. Web.
Abazov, V M, Budyashov, Yu G, Zorin, V P, Mitsyn, G V, Molokanov, A G, Savchenko, O V, & Astrakhan, B V. The approbation of rotating scanning irradiation method at therapeutic proton beam of the Laboratory of Nuclear Problems, JINR; Aprobatsiya metoda rotatsionnogo skaniruyushchego oblucheniya na meditsinskom protonnom puchke Laboratorii yadernykh problem, OIYaI. JINR.
Abazov, V M, Budyashov, Yu G, Zorin, V P, Mitsyn, G V, Molokanov, A G, Savchenko, O V, and Astrakhan, B V. 1993. "The approbation of rotating scanning irradiation method at therapeutic proton beam of the Laboratory of Nuclear Problems, JINR; Aprobatsiya metoda rotatsionnogo skaniruyushchego oblucheniya na meditsinskom protonnom puchke Laboratorii yadernykh problem, OIYaI." JINR.
@misc{etde_10144755,
title = {The approbation of rotating scanning irradiation method at therapeutic proton beam of the Laboratory of Nuclear Problems, JINR; Aprobatsiya metoda rotatsionnogo skaniruyushchego oblucheniya na meditsinskom protonnom puchke Laboratorii yadernykh problem, OIYaI}
author = {Abazov, V M, Budyashov, Yu G, Zorin, V P, Mitsyn, G V, Molokanov, A G, Savchenko, O V, and Astrakhan, B V}
abstractNote = {The method of rotating scanning irradiation with a proton beam for deeply located tumours is described. The Bragg peak and the tumour overlapping are controlled by a computer on the basis of information from miniature silicon detector inserted in the cavity inside the patient body near the tumour or from calculations from X-ray tomographic images, measured immediately before irradiation runs. In phantom tests good agreement between calculations from tomographic images and measurements with silicon detector has been obtained. The method may be used in clinical research. (author). 13 refs.; 12 figs.}
place = {JINR}
year = {1993}
month = {Dec}
}