You need JavaScript to view this

Reassessment of high and intermediate doses by the residual dose and the PTTL

Abstract

LiF:Ti,MG TLD cards (G-1 and G-7) were irradiated to high and intermediate doses, in the range of 0.15-7.5 Gy and the residual dose and the PTTL (Photo-Transferred Thermoluminescence) were measures. some dosimetric properties (e.g., accuracy and fading) of the residual dose and PTTL were investigated. We measure the minimum reassessed exposure, as well, when using both methods. We found that the minimum detectable exposure is 13 mGy measured by the PTTL, and 360 mGy measured by residual dose. (authors).
Publication Date:
Jan 01, 1993
Product Type:
Technical Report
Report Number:
NRCN(TN)-142
Reference Number:
SCA: 560101; PA: AIX-24:029706; SN: 93000955720
Resource Relation:
Other Information: PBD: Jan 1993
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; THERMOLUMINESCENT DOSEMETERS; ACCURACY; LITHIUM FLUORIDES; DOSE RATES; RADIATION DOSES; THERMOLUMINESCENT DOSIMETRY; 560101; DOSIMETRY AND MONITORING
OSTI ID:
10134764
Research Organizations:
Israel Atomic Energy Commission, Beersheba (Israel). Nuclear Research Center-Negev
Country of Origin:
Israel
Language:
Hebrew
Other Identifying Numbers:
Other: ON: DE93619305; TRN: IL9304945029706
Availability:
OSTI; NTIS (US Sales Only); INIS
Submitting Site:
INIS
Size:
[15] p.
Announcement Date:
Jul 04, 2005

Citation Formats

Ben-Shachar, Barak, German, Uzi, Rachmian, Ytzchak, and Ephraim, Naim. Reassessment of high and intermediate doses by the residual dose and the PTTL. Israel: N. p., 1993. Web.
Ben-Shachar, Barak, German, Uzi, Rachmian, Ytzchak, & Ephraim, Naim. Reassessment of high and intermediate doses by the residual dose and the PTTL. Israel.
Ben-Shachar, Barak, German, Uzi, Rachmian, Ytzchak, and Ephraim, Naim. 1993. "Reassessment of high and intermediate doses by the residual dose and the PTTL." Israel.
@misc{etde_10134764,
title = {Reassessment of high and intermediate doses by the residual dose and the PTTL}
author = {Ben-Shachar, Barak, German, Uzi, Rachmian, Ytzchak, and Ephraim, Naim}
abstractNote = {LiF:Ti,MG TLD cards (G-1 and G-7) were irradiated to high and intermediate doses, in the range of 0.15-7.5 Gy and the residual dose and the PTTL (Photo-Transferred Thermoluminescence) were measures. some dosimetric properties (e.g., accuracy and fading) of the residual dose and PTTL were investigated. We measure the minimum reassessed exposure, as well, when using both methods. We found that the minimum detectable exposure is 13 mGy measured by the PTTL, and 360 mGy measured by residual dose. (authors).}
place = {Israel}
year = {1993}
month = {Jan}
}