Abstract
Specifications for the maximum level of vibrational amplitudes at the sample position of the IRI neutron reflectometer ROG are presented. The acceleration and displacement amplitudes in the reactor floor have been measured as a function of frequency. These measured values meet the ROG specifications. (author). 11 figs.; 1 tab.
Citation Formats
Well, A.A. van.
Neutron reflectometry with ROG and mechanical vibrations.
Netherlands: N. p.,
1991.
Web.
Well, A.A. van.
Neutron reflectometry with ROG and mechanical vibrations.
Netherlands.
Well, A.A. van.
1991.
"Neutron reflectometry with ROG and mechanical vibrations."
Netherlands.
@misc{etde_10129463,
title = {Neutron reflectometry with ROG and mechanical vibrations}
author = {Well, A.A. van}
abstractNote = {Specifications for the maximum level of vibrational amplitudes at the sample position of the IRI neutron reflectometer ROG are presented. The acceleration and displacement amplitudes in the reactor floor have been measured as a function of frequency. These measured values meet the ROG specifications. (author). 11 figs.; 1 tab.}
place = {Netherlands}
year = {1991}
month = {Nov}
}
title = {Neutron reflectometry with ROG and mechanical vibrations}
author = {Well, A.A. van}
abstractNote = {Specifications for the maximum level of vibrational amplitudes at the sample position of the IRI neutron reflectometer ROG are presented. The acceleration and displacement amplitudes in the reactor floor have been measured as a function of frequency. These measured values meet the ROG specifications. (author). 11 figs.; 1 tab.}
place = {Netherlands}
year = {1991}
month = {Nov}
}