Abstract
Fifteen Spanish institutions took part in the European intercomparison of dosimeters promoted by DGXI of the EEC and steered by the German National Laboratory, PTB. The aim was to ascertain the measurement errors in the assessment of air kerma for the energy ranges corresponding to conventional diagnostic and mammography. Technical aspects are pointed out and the scheme proposed by PTB is described, together with the technical peculiarities of the Spanish branch, coordinated by the metrology division of CIEMAT. The results of the Spanish participation are presented and compared with the whole of the European participants. Some insight was also obtained on the performance of the measurement equipment and on the influence of some factors upon the air kerma measurements. (author) 27 fig. 2 ref.
Citation Formats
Brosed, A, Granados, C E, and Delgado, A.
The spanish participation in the European intercomparison of instruments used for dosimetry in diagnostic radiology; La participacion espanola en la intercomparacion europea de equipos de medida usados en dosimetria de radiodiagnostico.
Spain: N. p.,
1991.
Web.
Brosed, A, Granados, C E, & Delgado, A.
The spanish participation in the European intercomparison of instruments used for dosimetry in diagnostic radiology; La participacion espanola en la intercomparacion europea de equipos de medida usados en dosimetria de radiodiagnostico.
Spain.
Brosed, A, Granados, C E, and Delgado, A.
1991.
"The spanish participation in the European intercomparison of instruments used for dosimetry in diagnostic radiology; La participacion espanola en la intercomparacion europea de equipos de medida usados en dosimetria de radiodiagnostico."
Spain.
@misc{etde_10121280,
title = {The spanish participation in the European intercomparison of instruments used for dosimetry in diagnostic radiology; La participacion espanola en la intercomparacion europea de equipos de medida usados en dosimetria de radiodiagnostico}
author = {Brosed, A, Granados, C E, and Delgado, A}
abstractNote = {Fifteen Spanish institutions took part in the European intercomparison of dosimeters promoted by DGXI of the EEC and steered by the German National Laboratory, PTB. The aim was to ascertain the measurement errors in the assessment of air kerma for the energy ranges corresponding to conventional diagnostic and mammography. Technical aspects are pointed out and the scheme proposed by PTB is described, together with the technical peculiarities of the Spanish branch, coordinated by the metrology division of CIEMAT. The results of the Spanish participation are presented and compared with the whole of the European participants. Some insight was also obtained on the performance of the measurement equipment and on the influence of some factors upon the air kerma measurements. (author) 27 fig. 2 ref.}
place = {Spain}
year = {1991}
month = {Dec}
}
title = {The spanish participation in the European intercomparison of instruments used for dosimetry in diagnostic radiology; La participacion espanola en la intercomparacion europea de equipos de medida usados en dosimetria de radiodiagnostico}
author = {Brosed, A, Granados, C E, and Delgado, A}
abstractNote = {Fifteen Spanish institutions took part in the European intercomparison of dosimeters promoted by DGXI of the EEC and steered by the German National Laboratory, PTB. The aim was to ascertain the measurement errors in the assessment of air kerma for the energy ranges corresponding to conventional diagnostic and mammography. Technical aspects are pointed out and the scheme proposed by PTB is described, together with the technical peculiarities of the Spanish branch, coordinated by the metrology division of CIEMAT. The results of the Spanish participation are presented and compared with the whole of the European participants. Some insight was also obtained on the performance of the measurement equipment and on the influence of some factors upon the air kerma measurements. (author) 27 fig. 2 ref.}
place = {Spain}
year = {1991}
month = {Dec}
}