You need JavaScript to view this

Magnetic spectrometer of the DEUTERON-2 set-up

Abstract

A magnetic spectrometer of the two-arm DEUTERON-2 set-up of the Erevan Physical Institute is described. It is shown that the rejection factor for electrons and pions is 10{sup -2} - 10{sup -3}. The positively charged particles in the momentum range up to 1.5 GeV/c are identified by momentum and time-of-flight measurements. The main characteristics of the spectrometer are: momentum and angular acceptance {delta}p/p = 46%, {Delta}{theta} = 4 deg, solid angle {Delta}{Omega} = 2.75 msr, momentum resolution {delta}p/p = 1.5%, angular resolutions {delta}{theta} = 0.6 deg, {delta}{phi} = 2 deg. The intervals of measured momentum and the polar scattering anlge are 0.5-3 GeV/c and 10-30 deg, 68-90 deg respectively. 7 refs.; 11 figs.
Authors:
Ajvazyan, R V; Alanakyan, K V; Amaryan, M J [1] 
  1. and others
Publication Date:
Dec 31, 1989
Product Type:
Technical Report
Report Number:
YERPHI-1155-32-89
Reference Number:
SCA: 440104; 430303; PA: AIX-23:017418; SN: 92000661270
Resource Relation:
Other Information: PBD: 1989
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 43 PARTICLE ACCELERATORS; MAGNETIC SPECTROMETERS; PARTICLE DISCRIMINATION; CALIBRATION; CHARGED PARTICLE DETECTION; EFFICIENCY; EREVAN SYNCHROTRON; FLOWSHEETS; LINEAR MOMENTUM RESOLUTION; SPATIAL RESOLUTION; SPECIFICATIONS; 440104; 430303; HIGH ENERGY PHYSICS INSTRUMENTATION; EXPERIMENTAL FACILITIES AND EQUIPMENT
OSTI ID:
10117649
Research Organizations:
Erevanskij Fizicheskij Inst., Erevan (Armenia)
Country of Origin:
USSR
Language:
English
Other Identifying Numbers:
Other: ON: DE92616775; TRN: SU9110181017418
Availability:
OSTI; NTIS (US Sales Only); INIS
Submitting Site:
INIS
Size:
22 p.
Announcement Date:
Jun 30, 2005

Citation Formats

Ajvazyan, R V, Alanakyan, K V, and Amaryan, M J. Magnetic spectrometer of the DEUTERON-2 set-up. USSR: N. p., 1989. Web.
Ajvazyan, R V, Alanakyan, K V, & Amaryan, M J. Magnetic spectrometer of the DEUTERON-2 set-up. USSR.
Ajvazyan, R V, Alanakyan, K V, and Amaryan, M J. 1989. "Magnetic spectrometer of the DEUTERON-2 set-up." USSR.
@misc{etde_10117649,
title = {Magnetic spectrometer of the DEUTERON-2 set-up}
author = {Ajvazyan, R V, Alanakyan, K V, and Amaryan, M J}
abstractNote = {A magnetic spectrometer of the two-arm DEUTERON-2 set-up of the Erevan Physical Institute is described. It is shown that the rejection factor for electrons and pions is 10{sup -2} - 10{sup -3}. The positively charged particles in the momentum range up to 1.5 GeV/c are identified by momentum and time-of-flight measurements. The main characteristics of the spectrometer are: momentum and angular acceptance {delta}p/p = 46%, {Delta}{theta} = 4 deg, solid angle {Delta}{Omega} = 2.75 msr, momentum resolution {delta}p/p = 1.5%, angular resolutions {delta}{theta} = 0.6 deg, {delta}{phi} = 2 deg. The intervals of measured momentum and the polar scattering anlge are 0.5-3 GeV/c and 10-30 deg, 68-90 deg respectively. 7 refs.; 11 figs.}
place = {USSR}
year = {1989}
month = {Dec}
}