Abstract
The aim of the paper is calibration of units of a multichannel analyzer of charge-exchange neutrals intended for hot plasma diagnoctics. A dependence is obtained between the microchannel-plate detector output current and fast H{sup 0} current at the analyzer input in the 5-80 keV range. Operation modes of a gas target and the amplification coefficient of the microchannel-plate detector are determined. 9 refs., 9 figs.
Citation Formats
Antipin, D V, and Churkin, I N.
Calibration of multichannel analyzer of charge-exchange atoms; Kalibrovka mnogokanal`nogo analizatora atomov perezaryadki.
Russian Federation: N. p.,
1994.
Web.
Antipin, D V, & Churkin, I N.
Calibration of multichannel analyzer of charge-exchange atoms; Kalibrovka mnogokanal`nogo analizatora atomov perezaryadki.
Russian Federation.
Antipin, D V, and Churkin, I N.
1994.
"Calibration of multichannel analyzer of charge-exchange atoms; Kalibrovka mnogokanal`nogo analizatora atomov perezaryadki."
Russian Federation.
@misc{etde_10112840,
title = {Calibration of multichannel analyzer of charge-exchange atoms; Kalibrovka mnogokanal`nogo analizatora atomov perezaryadki}
author = {Antipin, D V, and Churkin, I N}
abstractNote = {The aim of the paper is calibration of units of a multichannel analyzer of charge-exchange neutrals intended for hot plasma diagnoctics. A dependence is obtained between the microchannel-plate detector output current and fast H{sup 0} current at the analyzer input in the 5-80 keV range. Operation modes of a gas target and the amplification coefficient of the microchannel-plate detector are determined. 9 refs., 9 figs.}
place = {Russian Federation}
year = {1994}
month = {Dec}
}
title = {Calibration of multichannel analyzer of charge-exchange atoms; Kalibrovka mnogokanal`nogo analizatora atomov perezaryadki}
author = {Antipin, D V, and Churkin, I N}
abstractNote = {The aim of the paper is calibration of units of a multichannel analyzer of charge-exchange neutrals intended for hot plasma diagnoctics. A dependence is obtained between the microchannel-plate detector output current and fast H{sup 0} current at the analyzer input in the 5-80 keV range. Operation modes of a gas target and the amplification coefficient of the microchannel-plate detector are determined. 9 refs., 9 figs.}
place = {Russian Federation}
year = {1994}
month = {Dec}
}